Start-up                                                      

   

1  .   .   .   .   .   .   .   .   .   Turn ON Master Switch

2  .   .   .   .   .   .   .   .   .   .   . Log on to Computer
3  .   .   .   .   .   .   .   .   .   .   .   .   .   Load Software
Optional   .   .   .   .   .Turn ON X-Cite Hg Lamp

Align Optical Components                 

1  .   .   .   .   .   .   .   .   .   .   .Koehler Illumination

Calibrate System                                    

1  .   .   .   .   .   .   .   .   .   .   .   .   .   Grid Tune Lens
2  .   .   .   .   .   . Parcentric Parfocal Calibration

Run Experiment                                       

1  .   .   .   .   .   .   .   .   .   .   .   .   .   .Set up Subject
2  .   .   .   .   .   .   .   .   .   .  Set scope to Low Mag
3  .   .   .   .   .   .   .   .  Trace Region(s) of interest
4  .   .   .   .   .   .   .   .   .   .Set scope to High Mag
5  .   .   .   .   .   .   .  Define Counting Frame Size
6  .   .   .   .   .   .   .   .   .   Define SRS Grid Layout
7  .   .   .   .   .   .   .   .   . Define Disector Options
8  .   .   .   .   .   .   .   .   Save Sampling Paramters
9  .   .   .   .   .   .   .  Select Markers for Counting
10   .   .   .   .   .   .   .   .   .   .   .   .   .Count Objects
11    .   .   .   .   .   .   .   .   .View Sampling Results

Save Files

 .   .   .   .   .   .   .   .   .   .   .   .   .  

1

  .   .   .   .   .Save to local machine-(often)

2  .   .   .   .   .   .   .   Copy Files Server (Zeus)

1.  Koehler align optic train.
2.  Deflect light to camera with push 

pull rod located above binocular vertical 

control knob.
3.  Start StereoInvestigator. Select Imag-

ing > Live Image (Ctrl-L)
4.  Click with mouse inside image frame 

to set reference point and to improve 

performance of mouse when hovering 

over the display area.
5.  Select Tools > Video > Camera Set-

tings... to bring up Camera Settings win-

dow to help with setting exposure Gain 

and other camera specific options.

6.  Adjust Exposure and white balance 

using More... button in the Camera Set-

tings window, another window will pop 

up  with the basic Olympus PictureFrame 

software interface “Auto exposure” and  

“AWB” auto white balance are the two 

most important functions. Ensure to 

select an area for white balance in the 

Olympus PictureFrame “live window.”

With StereoInvestigator

Acquire Images

7. Create a flatfield (optional)
Find a clear area for bright field, or even-

ly fluorescent field for epi-fluorescence; 

fluorescence dye slides are available on 

request.
  Capture image: Imaging > Acquire 

Image. Set the captured image as a 

background image:
  Flatfield for Brightfield: Imaging >   

Set to Brightfield Background Image
 

or

  Flatfield for Fluorescent: Imaging > 

Set to Fluorescent Background image
Tell software to subtract background 

image during acquisition: Imaging > En-

able Background Correction.
8. Acquire images: Imaging > Acquire 

Image. The acquired image will replace 

the live image remember to select Imag-

ing > Live Image (Ctrl+L) to return to live 

viewing of sample.

9.  Save Images select File > Image Save 

or Image Save As to save captured im-

ages.

Содержание AxioImager A1

Страница 1: ...Instructions for Zeiss AxioImager A1 Microscope for Stereology Basic operation instructions as of 10 1 2009 The Stereology Workstation By Jon Ekman...

Страница 2: ...fluorescence viewing Press the RL Key Set Reflector Turret to 2 Eyes DUAL 470EX 560EX 515EM 620EM Adjust the Incident Iris Stop Slider until light intensity is adequate Find Focus preferably on a dye...

Страница 3: ...crews 3 5 Open luminous field diaphragm 1 to edge of field 6 Adjust contrast using condenser aperture diaphragm slider 4 In most cases adjust aperture diaphragm slider to mark matching the NA of the o...

Страница 4: ...oscope and in the software Align for Koehler Illumination Grid Tune Calibration 8 Select Tools Grid Tune Current Lens 9 Tell the software the box size Large grid 250mm Small grid 25mm 10 Drop the anch...

Страница 5: ...o ing a pilot study will ensure that quality data is generated by obtaining the best estimates of total cell number Step 1 Set up the subject 1 Number of sections to count De fault is 1 new sections c...

Страница 6: ...eing counted NOTE Section Thickness measurements Guard Zones Step 8 Define Dissector Options Set Guard zones so that an over or under estimation of cell totals does not occur Do not count cells that f...

Страница 7: ...of Results The first items to appear in the results window are the parameters for your study Marker will show the estimated total cell numbers Do not be alarmed if some of the results are 0 Here is t...

Страница 8: ...with setting exposure Gain and other camera specific options 6 Adjust Exposure and white balance using More button in the Camera Set tings window another window will pop up with the basic Olympus Pic...

Страница 9: ...3 Plan Apochromat excellent quality cor rect for spherical aberration and corrected for chromatic aberration in red blue and green wavelengths 4 EC Epiplan Enhanced Contrast Univer sal objectives for...

Страница 10: ...as the Z drive in the windows environment ITG is equipped with gigabit ethernet and saving to the Z drive is relatively fast Also this permits easy access to the data from any computer in the world u...

Страница 11: ...hers 1998 Michalet X Kapanidis A N Laurence T Pinaud F Doose S Pflughoefft M and Weiss S The power and prospects of fluorescence microscopies and spectrosco pies Annu Rev Biophys Biomolec Struct 32 16...

Отзывы: