INSTALLATION INSTRUCTIONS AND FIRST-TIME SET-UP
ZEISS
Condensers
Axio Observer
52
431004-7244-001
12/2016
4.10.2
Fitting condensers from the
Axio Imager range
The following condensers from the Axio Imager
range can be used:
−
Condenser, LD achromatic 0.8 H DIC (424206-
9901-000)
−
Achromatic aplanatic condenser 1.4 H D Ph DIC
(424208-0000-000)
The Axio Observer's inverted design
means that these condensers must be
fitted "backwards" (with the turret at
the rear), such that the controls are at
the back and the labeling is upside
down.
•
Tilt the carrier for transmitted-light illumination
(Fig. 28/
4
) backwards and rotate the condenser
carrier to the highest position.
•
Insert the condenser mount (Fig. 32/
1
) into the condenser carrier on the carrier for transmitted-light
illumination with the dovetail ring facing upwards. Ensure that the locating pin on the condenser is at
the front and engages precisely with the guide groove on the condenser carrier.
•
Fix the condenser mount in place using the clamping screw (Fig. 32/
2
).
•
Insert the required condenser into the dovetail ring of the condenser mount, checking that it is
correctly orientated, and fix into position using the clamping screw (Fig. 32/
3
).
•
Carefully tilt the carrier for transmitted-light illumination (Fig. 28/
4
) back into the forward position.
Ensure that the condenser does not hit the specimen or the mounting frame.
•
Set the illumination to "KÖHLER".
Condensers from the Axio Imager range may only be used for specimen on slides.
It is not possible to combine them with the mounting frame K for reflected light (451344-
0000-000) or the objective piezo focusing units.
When using the achromatic-aplanatic condenser 1.4 it is recommended to use only mounting
frames with clamping springs for clamping the specimen slides.
Fig. 32
Fitting the condenser mount