Microscopy Core Facility 

 
 

 

V1.0 

 

It is possible to choose various methods of acquisition of an image and most 

actions capable from the software are possible. This will show you how to collect a 
Z-Stack. If you only desire a single multichannel image you can leave out to portions 
about Z-stack. Remember to not select a Z-stack (section f) ) 

 
Fig 4 – Axiovision Tool Bar 
 
 

 
 
 
 
 
a)

 

Select Workarea tool bar if workarea is not showing – Fig 4 

 

Fig 5 – Work area  
 
 

b)    

Select  Objective  of  choice  (ensuring 
oil  is  used  as  previously  described 
above  in  2c)  or  leave  as  presently 
chosen - Fig 5 

 

c) 

Select ‘Multidimensional Acquisition’ 
sub tab (as shown also in Fig 6) 

 

Содержание Axio Imager A1

Страница 1: ...sic coverage of use of the Zeiss Axio Imager M1 for multi channelled fluorescent acquisition For more complex usages please contact Fig 1 Front View of Microscope Ocular Viewer Main Stage Mercury Lamp...

Страница 2: ...mp control and Microscope power boxes located proximally to main microscope stage Turn on Microscope switch located behind 1st focusing wheel b Turn on computer and Activate AxioVision Program Fig 2 S...

Страница 3: ...ieved by pressing stage dropping button and adding a tear sized drop on top of slide so cover slips are necessary and returning stage by pressing stage returning button Make sure at this point to pres...

Страница 4: ...u only desire a single multichannel image you can leave out to portions about Z stack Remember to not select a Z stack section f Fig 4 Axiovision Tool Bar a Select Workarea tool bar if workarea is not...

Страница 5: ...hannels that will be examined by microscope if chosen For example in the Experiment box shown in Fig 6 the name shown is dapi gfp rhod phase This file will make the microscope acquire images through a...

Страница 6: ...1 is intended for phase contrast using an x10 objective which is not currently installed i It is now possible to select the optimal settings for each channel by clicking on each channel separately an...

Страница 7: ...begin m Focus to beneath sample and click on stop n Select a slice distance so that z slices are at desired distance from each other o Return Viewing slider to the original position pulled out p Click...

Страница 8: ...sample in every channel desired Fig 10 In this format the data is still raw and is best saved to be processed separately afterwards Fig 10 Z stack Window a You can turn each channel collected on and o...

Страница 9: ...ocessing option The function of deconvolution is to clean up images to remove positional and chromatic aberrations There are a choice of 3 algorithms which each function in their own way Simply click...

Страница 10: ...b select initial and final frame of view desired and enter into parameters section of workarea StartPos and EndPos Fig 12 c Select dimension of view default of ZY d Click Start to create new image Fig...

Страница 11: ...ures is not automatically chosen to be the right magnification When you choose to annotate with a scale bar a scaling tab needs to be opened This allows you to set the scale of the image b Select scal...

Страница 12: ...age and selecting properties It is in this window that the Attributes tab can be selected where properties of annotations can be changed This is particularly useful when annotation colour is not contr...