28
Calibrating the compensation data
AURA
®
handheld NIR - Rev. 2
Maintenance and Repair Manual | AURA
®
handheld NIR - Rev. 2 | EN_40_070_0030I
3
Place the device in the stand and apply the attachment for external
referencing
Insert the device
2
into the stand
1
and fit the attachment for external
referencing
3
onto the measuring window.
4
Insert the reference
Insert the relevant reference
4
into the centering aid.
Make sure that the reference is correctly aligned with the attachment
(the nose of the reference should slot into the groove of the attachment).
Fig. 24
3
2
1
4
Prepare device for calibration