31
Calibrating the compensation data
AURA
®
handheld NIR - Rev. 2
EN_40_070_0030I | AURA
®
handheld NIR - Rev. 2 | Maintenance and Repair Manual
7
Select certificates
Enter your name in the “User” input field.
Use the
buttons to select your reference certificates.
Click
to proceed to the next step.
8
Referencing – Calibration of white reference
Place the white reference on the centering aid. Make sure that the markings
on the reference and the centering aid are aligned.
Click
OK
to start the calibration.