4.8.2 Calibrate
It is necessary to calibrate the device from time to time. This is may be required particularly
if ambient conditions have been changed since the last calibration. It is recommended to
calibrate the device at least every 2 days.
This function will automatically calibrate all available
geometries against a calibration standard. The current
calibration values are shown on the entry window of a
calibration. These can be changed within the setup menu
(e.g. for calibration on a second standard). When pressing
the Calibration button the ZGM 1130 moves on with the
calibration. If the device is not placed on the calibration
standard a message is shown on the display.
After the device has performed a calibration, a summary will
be shown indicating which angles exceeding a certain tolerance.
Confirm the calibration with
or discard it with
.
This angle has been calibrated successfully without any error.
This angle has been calibrated successfully without any error but the device
has not been placed on its calibration standard.
An error occurred during calibration. Please consider the message shown.
4.8.3 Menu
The menu items gives the user access to all settings of the device.