ZEGGO THA_C1Z1 

REV. 1 03/19 

 

 

Configuration (Admin) 

 

 

 

 

 

 

    

 

These are the configurations that could be accessed by Zeggo’s technician only.  

 

Reset Memory/ Time, Reset Battery 

This feature enables the zero calibrate the ADC, clear the internal storage and reset stored time in the 
device memory.  

i) Unplug the sensor.  

ii) Turn the slide switch in OFF position. 

iii) Press and hold the reset button on the PCB Board. 

iv) Turn the slide switch in ON Position.  

v) Wait for a preset of time and observe the status LED according to Table. 

vi) Release the reset button. 

 

Table 3: Configurations by Admin 

Time Holding Reset Button 

Status LED Indication 

Configuration Made 

5 Secs 

Green LED Lighted Up 

Zero Calibration 

15 secs 

Alternate Green Red (Slow) 

Clear Internal Memory, Reset 

Time 

30 secs 

Alternate Green Red (Fast) 

Reset Battery Counter/ 

Percentage 

 

Software Calibration 

Device will enter configuration mode from special sequence button press and hold.  

i) Make sure the device is in OFF state and wait for about 4 second.   

ii) Press the ON/OFF/configuration button for 7 times (For negative calibration)/ 9 times (For positive 
calibration).  (Interval must less than 2 secs) 

iii) Press and hold the button for the next attempt. 

The status LED will blink  GREEN and RED alternately upon successful enter configuration mode. If the 
status LED blink GREEN LED only, it fails to enter configuration but goes into normal ON mode. Kindly 
follow the steps again for another attempt. 

During configuration mode, 10 secs timeout is implemented if no input is received. Each press and release 
of button will go a step in the sending interval setting as below: 

Содержание THA C1Z1

Страница 1: ...ZEGGO THA_C1Z1 REV 1 03 19 1 Future Integrated Data Evaluation Wireless Humidity and Temperature Transmitter Model THA_C1Z1 USER MANUAL...

Страница 2: ...ZEGGO THA_C1Z1 REV 1 03 19 2 DOCUMENT HISTORY Release No Date Revision Description 1 28 03 19 First Release Humidity Transmitter THA_C1Z1 V1 0 2 16 04 19 System manual updated...

Страница 3: ...15 4 Radio Transmission to Zeggo s Gateway Easy Configuration and Installation 2 Years of Battery Life Internal storage to store during network failure Programmable sampling interval 1min 5min 10min...

Страница 4: ...elivery It is always a good practice to check that the device is ALIVE prior to the installation by turn ON OFF the device By pressing and holding the button for an interval blinking and colour of sta...

Страница 5: ...OFF configuration button for 4 times Interval must less than 2 secs iii Press and hold the button for the 5th attempt The status LED will blink GREEN and RED alternately upon successful enter configur...

Страница 6: ...ate Green Red Slow Clear Internal Memory Reset Time 30 secs Alternate Green Red Fast Reset Battery Counter Percentage Software Calibration Device will enter configuration mode from special sequence bu...

Страница 7: ...d ON Internal Memory Backup In the event when signal failed to send to a gateway the temperature data stores in the internal memory These backup data will send again to gateway after network recovery...

Страница 8: ...ommended to replace the battery if battery level is less than 20 The suggested signal percentage for a stable connection is 40 to take care the moving object or addition of object in between signal pa...

Страница 9: ...OF REVENUE OR PROFIT FAILURE OF OTHER EQUIPMENT TO WHICH PRODUCT IS CONNECTED WITH RESULTING FROM THE SUE OF PRODUCT RELATING TO WARRANTY SERVICE OR ARISING OUT OF ANY BEACH OF THIS LIMITED WARRANTY E...

Отзывы: