Configuration (Admin)
These are the configurations that could be accessed by Zeggo’s technician only.
Reset Memory/ Time, Reset Battery
This feature enables the zero calibrate the ADC, clear the internal storage and reset stored time in the
device memory.
i) Unplug the sensor.
ii) Turn the slide switch in OFF position.
iii) Press and hold the reset button on the PCB Board.
iv) Turn the slide switch in ON Position.
v) Wait for a preset of time and observe the status LED according to Table.
vi) Release the reset button.
Table 3: Configurations by Admin
Time Holding Reset Button
Status LED Indication
Configuration Made
5 Secs
Green LED Lighted Up
Zero Calibration
15 secs
Alternate Green Red (Slow)
Clear Internal Memory, Reset
Time
30 secs
Alternate Green Red (Fast)
Reset Battery Counter/
Percentage
Software Calibration
Device will enter configuration mode from special sequence button press and hold.
i) Make sure the device is in OFF state and wait for about 4 second.
ii) Press the ON/OFF/configuration button for 7 times (For negative calibration)/ 9 times (For positive
calibration). (Interval must less than 2 secs)
iii) Press and hold the button for the next attempt.
The status LED will blink GREEN and RED alternately upon successful enter configuration mode. If the
status LED blink GREEN LED only, it fails to enter configuration but goes into normal ON mode. Kindly
follow the steps again for another attempt.
During configuration mode, 10 secs timeout is implemented if no input is received. Each press and release
of button will go a step in the sending interval setting as below: