background image

 

 

 

 

 

 

DSO3D12

 

User’s Manual

 

Designed by ZEEWEII     

 

DIGITAL OSCILLOSCOPE 

Содержание DSO3D12

Страница 1: ...DSO3D12 User s Manual Designed by ZEEWEII DIGITAL OSCILLOSCOPE...

Страница 2: ...de is normal Trigger edge Rising edge trigger or falling edge trigger Battery level indication Trigger level position X1 X10 indicates probe attenuation which should be set to match the probe The goal...

Страница 3: ...eform 50 Quick centering CH1 DMM Turn the multimeter on off Press Shift DMM to full screen Select Multimeter function selection mV V Adjust vertical sensitivity voltage of one grid s ns Adjust the tim...

Страница 4: ...connected the oscilloscope automatically adjusts the range in real time Press any key to exit Shift V ch1 Turn on off the voice assistant 1 Safety Precautions High voltage The x10 range of the probe...

Страница 5: ...anual adjustment mV V buttons scale the waveform vertically The s ns button scales the waveform horizontally The vertical sensitivity indicates the voltage of a grid in the vertical direction and it i...

Страница 6: ...set the high priority of channel 2 pay attention to the change of the 1 2 icon in the upper left corner When not using dual channel measurements Channel 2 should be turned off for higher sample rates...

Страница 7: ...s 3 4 1 Large input capacitance Because the probe 1X has a large input capacitance usually more than 100 pf it is not suitable for measuring some capacitively sensitive circuits For example when measu...

Страница 8: ...f the signal by observing how many divisions the waveform occupies on the screen For example if the vertical sensitivity is 1v div and the signal occupies 3 grids on the screen you can know that the a...

Страница 9: ...for adjusting the time base in the oscilloscope button panel are ns and s button When observing high frequency signals you should press the ns button to decrease the time base When observing low freq...

Страница 10: ...The arrow circled above is the trigger level position and when the edge of the square wave in the above figure passes upwards through the position of this arrow a trigger is formed the waveform only...

Страница 11: ...ancel a single trigger click Stop to stop Procedure 1 Connect the signal first 2 mV V set the voltage sensitivity s ns set the time base 3 Set the trigger level to manual and adjust the trigger level...

Страница 12: ...exit can adjust the duty cycle other waveforms do not work When the set frequency is greater than 1MHz the square wave signal may show small horizontal jitter Single dual channel button to turn on off...

Страница 13: ...e 1X 8 2 Measurement There are 14 measurement options available as shown below Frequency Peak to Peak Duty cycle Amplitude RMS Average Period Pulse width Pulse width Max Min Top Base Duty cycle The di...

Страница 14: ...emain on the screen for a while and then disappear Optional persistence options are None 1s Infinite When Infinity is selected the waveforms that have appeared on the screen will remain forever and wi...

Страница 15: ...Support Equivalent sample 500M Screenshot Support Rise time 3ns Frequency 0 01 Storage depth 64K Single Support Impedance 1M FFT Support Time base 5ns 10s Wave out amp 2 5V 0 05 Ver sensitivity 10mV...

Страница 16: ...ice speech rate accent etc we cannot guarantee the accuracy of recognition Wake up command Hello ZEEWEII Pronounced Hello see wave or Hello mini scope Command words function Auto setting It is equival...

Страница 17: ...w do I reset the device to its initial state Answer Open Menu Set and then long press the Auto button to reset 1 What is the difference between probe 1X and 10X Answer The 1X does not attenuate the si...

Страница 18: ...cted to the audio output The time base is adjusted to about 1 10ms and the voltage sensitivity is adjusted to about 500mV decided according to the volume 6 How to measure main board communication sign...

Отзывы: