3.2

 

Vertical position 

The method  of moving the 

waveform

 up  and down: directly press the up  and 

down keys to move the waveform up and down. 

In addition

: By default, the up and down keys move the waveform. If you need 

to move the trigger level, you need to hold down the “Power” key and then click the 
up and down keys to move the trigger level. 

One-handed operation  (not  recommended). You  can  also  long  press the  "mV" 

buttons to switch between the "Waveform" or "Trigger Level". Pay attention to the 
change of the arrow in the upper left corner, as shown below:   

4  Horizontal system 

4.1  Time base 

The  time  base  refers  to  the  time  represented  by  each  grid  in  the  horizontal 

direction on the oscilloscope screen. There are 12 grids in the horizontal direction of 
the oscilloscope. The sampling time that can be observed on the screen = " time base" 
*  12.The  user  can  estimate  the  signal  period  by  observing  how  many  grids  a 
waveform period occupies. For example, the current time base is 500us, and one cycle 
of the signal occupies 2 grids,so the signal period is 1ms (1KHz).

 

 

The oscilloscope time base range is 5ns ~ 10s. The buttons for adjusting the time 

base  in  the  oscilloscope  button  panel  are  '

ns

'  and  '

s

'  button.  When  observing 

high-frequency  signals,  you  should  press  the  'ns'  button  to  decrease  the  time  base. 
When  observing  low-frequency  signals,  you  should  press  the  's'  button  Increase the 
base gear. 

What you need to know is that the sampling process of the oscilloscope is: 

"Sampling">>"Processing" >> "Display" 

When  adjusted  to  a  large  time  base,  the  screen  waveform  refresh  will  be  slow 

due to the longer time taken by the oscilloscope to sample the waveform. For example, 
if the user adjusts to the 50ms time base, that is, the entire sampling time is at least 
50msx12 = 600ms, the oscilloscope waveform refresh speed is reduced to less than 2 
frames of waveform per second, which is normal. 

4.2  Horizontally moving waveform 

  In the stop mode, the user can move the waveform left and right.   

Содержание DSO1511G

Страница 1: ...ZEEWEII DSO1511G User s Manual Designed by zeeweii DSO1511G HANDHELD OSCILLOSCOPE...

Страница 2: ...enerator When it is closed no display Time base which means the time of a grid in the horizontal direction Click the ns and s buttons to adjust Coupling DC AC Probe attenuation x1 x10 x100 Vertical se...

Страница 3: ...r buttons Power OK short press Press and hold Power and short press Ok to switch the trigger edge rising or falling Power OK long press Press and hold Power and long press Ok to switch the trigger mod...

Страница 4: ...VITY 7 3 2 VERTICAL POSITION 8 4 HORIZONTAL SYSTEM 8 4 1 TIME BASE 8 4 2 HORIZONTALLY MOVING WAVEFORM 8 5 TRIGGER SYSTEM 9 5 1 TRIGGER LEVEL 9 5 2 TRIGGER POSITION 9 6 OPERATION GUIDE 10 6 1 POWER 10...

Страница 5: ...danger be sure to use this product as specified Only authorized personnel should perform maintenance procedures Avoid fire and personal injury Use the probe correctly and ensure that the measured vol...

Страница 6: ...thout overshoot 3 Adjust the horizontal and vertical scales of the oscilloscope so that the waveform is displayed stably at a suitable position on the screen 4 Use a screwdriver to rotate the adjustab...

Страница 7: ...gh voltage signal otherwise the high voltage may break down the internal circuit of the oscilloscope 2 4 3 High bandwidth Because the 1X probe has a large input capacitance the bandwidth of the probe...

Страница 8: ...observing how many grids a waveform period occupies For example the current time base is 500us and one cycle of the signal occupies 2 grids so the signal period is 1ms 1KHz The oscilloscope time base...

Страница 9: ...In edge trigger mode a trigger condition occurs only when the waveform crosses the trigger level To adjust the trigger level 1 Change Trigger Level to Manual in the menu 2 Method 1 Press and hold the...

Страница 10: ...rging current is about 700ma The LED at the bottom of the charging process is bright red and the led turns green when fully charged 6 3 Vertical Time base scale Press the mV and V buttons to adjust th...

Страница 11: ...s sampling is not completed until there is a waveform trigger Operation method Click the xKey button and select Single 6 9 Reference waveform The reference waveform is that the waveform of the previou...

Страница 12: ...square wave signal may show small horizontal jitter 6 13 Video output The 3 5mm headphone jack at the bottom can output AV analog video signal which can be connected to a TV or projector Open Menu Set...

Страница 13: ...ltage of the dotted line below dY Y1 Y2 is the voltage difference between the two dotted lines 7 Menu introduction Click the Menu button to open close the menu 7 1 Channel The channel menu includes tw...

Страница 14: ...rigger mode is Normal you should use the manual option Under this option the trigger level can be manually adjusted by the user 7 4 Display In this menu you can set persistence roll mode and brightnes...

Страница 15: ...after no button operation 2 You can choose whether to turn on the sound 3 Language selection Chinese or English 4 In this menu long press the ns button to enter the usb connection state long press s...

Страница 16: ...16...

Отзывы: