zebris Medical GmbH
FDM Technical Data and Operating Instructions
Page 25/36
Synchronization input (SYNC-IN)
If a third party device is connected to the synchronization input SYNC-IN then depending
on the setting of the configuration window from the application software the measurement
will start/stop or “frame by frame” synchronized by a signal from the third party device.
Input is protected against faulty polarisation and pin 1 is set to +5V ("1") by an internal pull-
up- resistor 2.7 k
. If this input is set to 0 V ("0") i.e. by a switch or break contact than the
SYNC-IN is triggered.
Built-in LEMO
– Jack at Front of ZEBRIS Gauge
Series „00“, 3-pin, fitting Nut 30°
LEMO- Part. No. EPA.00.303.NLN
Respective Plug:
Type of plug for SYNC-IN: LEMO- Part No. FGA.00 303.CLADxxxx
Pin assignment:
Pin 1
Clk_In
Pin 2
Active_In
Pin 3
GND
Electrical specifications:
Input resistance (pull-up 5 V)
2.7 k
VIH (High-Level Input Voltage)
>= 2.0 V
VIL (Low-Level Input Voltage)
<= 0.8 V
Required min. pulse time for triggering 1 ms
View
Front side
View
Solder side
Fitting Nut: 30°
Fitting Nut: 30°