zebris FDM Series Скачать руководство пользователя страница 25

zebris Medical GmbH 

FDM Technical Data and Operating Instructions 

Page 24/36

 

Example configurations for the synchronization of third party devices 

 
 

zebris system is providing the sync-signal 

 
 

Synchronize external 3

rd

 party device by start/stop signal 

 
Enable 

“Active out”

. Connect the 

“Active Out“ pin of the SYNC-OUT socket

 to 

start/stop input pin of the other device. The active out is set when the measurement 
is started and removed when the measurement stops. If the external device needs 
its start pin to be pulled low, set active out polarity to active low. 
Note: Output pins may toggle during device initialization. 
 
 

Synchronize  external  3

rd

  party  device  sample-by-sample  via  clock  signal 

generated by the zebris system 

 
Enable 

“Clock  out”

.  Connect  the 

“Clock  Out“  Pin  of  the  SYNC-OUT  socket 

to 

the trigger  input  of the other  device. The device  will generate a trigger pulse  each 
time a single measurement is recorded. 

 
 

Third party device is providing the sync-signal 

 
 

Synchronize the zebris device by start/stop from an external 3

rd

 party device 

 
Enable 

“Active  in”

.  Connect  the  activation  signal  from  the  other  device  to  the 

Active 

In“ Pin from the SYNC-IN socket

 of the zebris device.  

If the external device sets its output to low level during measurement, set the active 
in polarity to “active low”. 
 
 

Synchronize the zebris device sample-by-sample by a clock generated by an 
external 3

rd

 party device 

 
Enable 

“Clock in”

. Connect the clock signal from the external device to the 

„Clock 

In“ Pin from the SYNC-IN socket

 of the zebris device.  

If  the  zebris device  should  trigger  its  measurement  at  the falling edge of the clock 
signal,  set  the  clock  in  polarity  to  “active  low”.  Otherwise,  the  raising  edge  will 
trigger the measurement. 

 
 

Combinations 

 

All  signals  can  be  used  simultaneously  as  well.  If  both  active  In  and  clock  In  are 
enabled, the clock signal will trigger measurements only while the active In signal is 
present. Trigger pulses from the clock In signal which occur while there is no active 
In signal will be ignored. 

 

Содержание FDM Series

Страница 1: ...zebris Medical GmbH FDM Text Release 02 07 2013 Measuring System for Gait and Stance Analysis FDM Specifications and Operating Instructions...

Страница 2: ...f the FDM platform 19 4 1 Synchronizing the FDM Platform with video data Sync Audio 19 4 2 Infrared synchronization with zebris DAB Bluetooth EMG 20 4 3 Combine two FDM platforms of the same type 21 4...

Страница 3: ...e of any errors when using this User Manual or should you find details that do not conform with your device please kindly inform us We shall then correct any possible errors as quickly as possible In...

Страница 4: ...d and how it can be avoided Notes are shown in the following way NOTE NOTE symbols indicate a potential danger that can cause damage to or destroy the device These NOTE symbols describe the danger inv...

Страница 5: ...s using the system later The patient must be cognitively able to follow the operator s instructions or otherwise has to be assisted by trained professionals The application and operation of the system...

Страница 6: ...ccurate measuring results If any fluids should penetrate the device it is mandatory for the device to undergo a technical safety test Damaged plug connections and leads are to be replaced by an author...

Страница 7: ...ures or measures endangering the patient be carried out solely on the basis of the measuring results without further verification of the measuring data by additional methods Prohibited Use We warn exp...

Страница 8: ...safety for the test person the operator and the surroundings by consulting the manufacturer the authorized zebris sales partner or by asking a specialist WARNING If the computer is not supplied with...

Страница 9: ...to the respective measuring column The analog signal coupled into the shift register over the lines is proportional to the pressure dependent capacity and is passed on for further processing to the c...

Страница 10: ...rfaces USB Synchronization input output Video synchronization Infrared synchronization Connection Interface box on the bottom of housing frame Measuring principle capacitive force measurement Operatin...

Страница 11: ...f FDM multi function force plates Platform Type FDM SX Item number 124 3010 Dimensions L x W x H 55 x 40 x 2 1 cm L x W x H Weight 4 8 kg Sampling frequency 120 Hz Number of sensor 1920 Sensor surface...

Страница 12: ...Platform Type FDM S Item number 124 3005 Dimensions L x W x H 69 x 40 x 2 1 cm L x W x H Weight 6 5 kg Sampling frequency 120 Hz optional 240 Hz Number of sensor 2560 Sensor surface 54 2 x 33 9 cm L x...

Страница 13: ...nce and gait analysis Platform Type FDM 1 5 Item number 124 3015 Dimensions L x W x H 158 x 60 5 x 2 1 cm L x W x H Weight approx 16 5 kg Sampling frequency 100 Hz optional 200 300 Hz Number of sensor...

Страница 14: ...rm Type FDM 2 Item number 124 3020 Dimensions L x W x H 212 x 60 5 x 2 1 cm L x W x H Weight approx 25 kg Sampling frequency 100 Hz optional 200 Hz Number of sensor 15360 Sensor surface 203 2 x 54 2 c...

Страница 15: ...6 Platform Type FDM 3 Item number 124 3030 Dimensions L x W x H 307 x 60 5 x 2 1 cm L x W x H Weight approx 35 kg Sampling frequency 100 Hz Number of sensor 22528 Sensor surface 298 1 x 54 2 cm L x W...

Страница 16: ...Pin arrangement polarity NOTE It is absolutely necessary to connect the external power supply unit to a power outlet that contains a ground contact in order to secure fault free and save operation of...

Страница 17: ...power supply unit is connected to mains and a correct supply voltage is provided A USB connection is established and recognized The platform is ready for initialization or measurement orange flashing...

Страница 18: ...NOTE Please make absolutely sure that you have installed the zebris software before connecting the FDM platform to the computer using the USB cable When this step is skipped it may result in severe p...

Страница 19: ...does not take any liability for any injury to persons damage to property or loss of data due to improper connection of the platform If the power supply unit and the USB cable are connected to the soc...

Страница 20: ...he FDM Platform with video data Sync Audio The Sync Audio socket serves for synchronizing the platform measurement and recordings of commercially available video cameras utilizing the external microph...

Страница 21: ...in the platform housing is required FDM platform and DAB Bluetooth are synchronized automatically as soon as both devices have been switched on and a measurement is started The following schematic dia...

Страница 22: ...No 1830019 SC PP Sync Cable length 10m Both platforms have to be connected to separate USB ports of the same PC By means of the synchronization cable the Sync Out socket of the master platform has to...

Страница 23: ...d by means of galvanic separation acc to to the provisions of IEC 601 1 when a third party device is synchronized with the FDM system This allows non medical equipment to be synchronized with the FDM...

Страница 24: ...then the system will start a measurement on each synchronization impulse Higher prescaler values work as divider In case the prescaler is set to 2 then a measurement will be started on every second im...

Страница 25: ...recorded Third party device is providing the sync signal Synchronize the zebris device by start stop from an external 3rd party device Enable Active in Connect the activation signal from the other de...

Страница 26: ...d pin 1 is set to 5V 1 by an internal pull up resistor 2 7 k If this input is set to 0 V 0 i e by a switch or break contact than the SYNC IN is triggered Built in LEMO Jack at Front of ZEBRIS Gauge Se...

Страница 27: ...of the third party device either via start stop or frame by frame mode Built in LEMO Jack at Front of ZEBRIS Gauge Series 00 4 pin fitting Nut 0 LEMO Part No EPG 00 304 NLN Respective Plug Type of plu...

Страница 28: ...measurement guaranties reproducible steps and results Notice These results differ more or less from these by normal walking 5 2 Data recording Please observe the exercise of the patient strictly Only...

Страница 29: ...f thereof must only be carried out by zebris Medical GmbH or a person who has been explicitly authorized by zebris to do this WARNING Should any malfunctions and or defects be determined or suspected...

Страница 30: ...m the surrounding values Whenever faults occur or in case of doubt the manufacturer or sales partner authorized by zebris must always be contacted Calibration measures The measuring accuracy of the se...

Страница 31: ...latform with a cloth soaked in disinfection liquid Recommended disinfection agent Composition approx 25 ethanol 35 Propanol E g Mikrozid Liquid Sch lke Mayr or similar agents NOTE If you apply disinfe...

Страница 32: ...of the domestic waste system Within Europe this device must be forwarded to a specific waste disposal system Therefore regular disposal is carried out by the manufacturer For this purpose the system s...

Страница 33: ...EN 60601 1 Electrical Safety Standard BF Safety class II Steady state conditions 8 3 Connecting the FDM System to other electrical devices Important notes on medical electronically systems NOTE You ar...

Страница 34: ...structions given below WARNING Even though the motion analysis system FDM fully complies with the requirements of the standard EN 60601 1 2 it cannot be completely exempted that portable and mobile RF...

Страница 35: ...kV contact discharge 8 kV atmospheric discharge 6 kV contact discharge 8 kV atmospheric discharge Flooring should be of wood or concrete or laid with ceramic tiles If the flooring is made of synthetic...

Страница 36: ...100 predictable A site study is recommended to determine the electromagnetic environment as a result of stationary RF transmitters If the measured field strength at the site of the FDM force distribu...

Страница 37: ...zebris Medical GmbH FDM Technical Data and Operating Instructions Page 36 36 8 5 Declaration of conformity medical platforms...

Страница 38: ...zebris Medical GmbH FDM Technical Data and Operating Instructions Page 37 36 8 6 Declaration of conformity non medical platforms...

Страница 39: ...zebris Medical GmbH FDM Technical Data and Operating Instructions Page 38 36 Notes...

Отзывы: