4.4 Synchronization of third party devices with the FDM system
The
SYNC-IN
and
SYNC-OUT
sockets provide inputs and outputs for support of „frame by
frame“ In- and Out synchronization. Both sockets are galvanic protected from the platform.
NOTE
Patient’s safety is guaranteed by means of galvanic separation acc.
to to the provisions of IEC 601-1 when a third party device is
synchronized with the FDM system. This allows non medical
equipment to be synchronized with the FDM system as long as such
devices are out of patients reach. Nevertheless the user is
completely responsible for the safety of all third party devices used
in combination with the FDM system.
The WinFDM software provides a configuration window which allows adapting the
synchronization exactly to the parameters of interfaces from third party devices.