Please read the provided operating and instruction manual
QUICK SET UP GUIDE ZANTINGH CO DETECTOR
Telephone (reception)
+31 (0)297 219 100
Telephone (service department) +31 (0)297 219 125
1.
Mounting the device
•
mount upright and level to a solid wall, do not lay flat (on a horizontal surface)
•
do not install without its lit, always leave the device covered
•
do not install in a humid environment
•
do not install close to a heat source or in a place at a temperature below zero degrees
•
never suspend from a condenser
or any other
vibrating
parts
2.
Connecting the hose
•
make sure the sampling hose
first goes upwards from the sampling point side
, to prevent condensate to flow back to
the sampling point
•
prevent that a water seal is formed (water trap / siphon effect)
•
make sure the angled connector with compression fittings is assembled correctly, the connector should point upwards
•
tighten the nuts well to fix the compression fittings (rotate 1¼ times)
•
the used sampling hose should never be
longer than 30 meters and not larger than Ø6mm
•
make sure that the following connections are always free:
-
condensate drain
-
sample gas out
-
by-pass out
-
calibration gas in