Please read the provided operating and instruction manual 

QUICK SET UP GUIDE ZANTINGH CO DETECTOR 

 

Telephone (reception) 

 

+31 (0)297 219 100

 

Telephone (service department)  +31 (0)297 219 125 

3.

 

Connecting the device electrically 

 

use the 6- and 7-pole connectors for input and output signals  

 

4.

 

Turning on the device 

 

after turning on the device, the CO detector is ready for service (metering) after 45 minutes 

 

adjust the needle valve on the flow meter for a correct by-pass flow (the bullet should be in the green section) 

 

never turn off the device 

 

 

 

 

 

 

5.

 

Functionality of the device 

 

the value on the display is the current measurement value of the CO gas concentration in ppm 

 

an error message 

OLL

 

or “OHH”

 is displayed when the measurement value is out of range (<0 ppm or >100 ppm) 

 

the display remains blanc if there is a temperature, flow or hardware issue 

 

while purging (once every 60 minutes) no CO gas should be present near the device; exhaust gasses and (cigarette) 
smoke have a negative effect on its functionality 

 

three bicolour LEDs indicate the status according to the table below 

 
 
 
 
 
 
 

 
 

 
 
 
 
 
 
 

6.

 

Maintenance 

 

the filter should be replaced every 6 months 

 

the device needs to be calibrated every 12 months; always use 100% nitrogen (N

2

) calibration gas

 

 

the sensor should be replaced every 2-3 years

 

LED name

RED

GREEN (constant)

Sampling

N/A

-

       

System is measuring.

-

     

Too little or no flow 

(flashes)

-

     

Internal temperature too low

(constantly on) 

-

     

CO concentration is too high.

(>1 minute at set value or at >70 
ppm) 

System failure

-

       

Flow and temperature are 

appropriate.

CO failure

-

       

CO concentration within 

programmed limits.