Please read the provided operating and instruction manual
QUICK SET UP GUIDE ZANTINGH CO DETECTOR
Telephone (reception)
+31 (0)297 219 100
Telephone (service department) +31 (0)297 219 125
3.
Connecting the device electrically
•
use the 6- and 7-pole connectors for input and output signals
4.
Turning on the device
•
after turning on the device, the CO detector is ready for service (metering) after 45 minutes
•
adjust the needle valve on the flow meter for a correct by-pass flow (the bullet should be in the green section)
•
never turn off the device
5.
Functionality of the device
•
the value on the display is the current measurement value of the CO gas concentration in ppm
•
an error message
“
OLL
”
or “OHH”
is displayed when the measurement value is out of range (<0 ppm or >100 ppm)
•
the display remains blanc if there is a temperature, flow or hardware issue
•
while purging (once every 60 minutes) no CO gas should be present near the device; exhaust gasses and (cigarette)
smoke have a negative effect on its functionality
•
three bicolour LEDs indicate the status according to the table below
6.
Maintenance
•
the filter should be replaced every 6 months
•
the device needs to be calibrated every 12 months; always use 100% nitrogen (N
2
) calibration gas
•
the sensor should be replaced every 2-3 years
LED name
RED
GREEN (constant)
Sampling
N/A
-
System is measuring.
-
Too little or no flow
(flashes)
-
Internal temperature too low
(constantly on)
-
CO concentration is too high.
(>1 minute at set value or at >70
ppm)
System failure
-
Flow and temperature are
appropriate.
CO failure
-
CO concentration within
programmed limits.