manualshive.com logo in svg
background image

32

Diagram #9:  DuraSite Portable Sample Vessel Instructions

Purpose:  The DuraSite Portable Sample Vessel permits the
user to remove a liquid or gas hydrocarbon sample from a
pipeline or a sampling device.  This is accomplished without
changing the pressure of the product or exposing it to a
contaminant fluid.  If properly used and maintained the
DuraSite will provide many years of safe, accurate and clean
sampling.

Use:  The DuraSite is a very safe device to use.  As with any
equipment dealing with flammable products, it is mandatory
that a good, thorough operator training procedure be
established prior to use.

Typical use of the cylinder would be as follows:

Step 1:  (In The Lab)  Connect a regulated inert gas supply to
the pre-charge valve.  The product valve should be open.  By
carefully controlling the
pre-charge valve and the regulator, the cylinder can be slowly
charged with pre-charge gas (NOTE:  This should be done
slowly to prevent slamming the piston down to the opposite
end).  The pressure on the pre-charge pressure gauge should
be brought to a reading of 10-50 psi above the expected
pressure of the product in the field .  Close the pre-charge valve
and disconnect the gas supply.  Check the pre-charge valve,
relief device, and the pre-charge pressure gauge for leaks.  Any
leaks should be stopped before continuing.  The vessel should
be placed in a padded carrying case and made ready for field
use.
Proceed to 

EITHER Step 2, or Step 3 as

required for your application.

STEP 2: FOR COLLECTION OF SAMPLE VIA SPOT SAMPLE  OR
FROM  COMPOSITE  ACCUMULATOR  VESSEL.

2a:  Connect the product end of the pre-charged sample vessel
to the product supply.
(Sampler product removal valve, or Pipeline sample probe)
NOTE:  the pre-charge pressure gauge reading should be
greater than the product supply pressure reading.  If not,
repeat Step 1 above.

2b:  Once the vessel is connected to the product supply, it is
necessary to vent a small amount of product prior to filling the
vessel.  This assures fresh product and removes any air or gas
when dealing with liquids.  This can be done by loosening the
product purge valve a very small amount until the product is
purged.  After thorough purging, the product purge valve should
be tightened.

2c: The product pressure gauge reading should be 10-50 psi
below the pre-charge pressure gauge reading. By carefully
opening the pre-charge valve, the pressure becomes equal-
ized, then begins to drop below the product pressure.  The pre-
charge valve should be carefully controlled so as to not vent the
pre-charge gas too fast.

2d:  When the cylinder becomes a maximum of 80% full (see
volume indicator), all valves should be closed.  The product
connection is slowly broken in order to vent any trapped product.
After vessel removal, all connections should be checked for
leaks and the pre-charge and product valve ports capped to
prevent leakage.

2e:  Pack the DuraSite in appropriate carrying case to meet
D.O.T. guideline, with D.O.T.
paperwork and transport to lab for analysis.

STEP 3: FOR DIRECT CONNECTION TO
SAMPLER.

3a:  Connect the sampler discharge port to the product inlet port
to the DuraSite using 1/8" stainless steel tubing.

3b:   (Gas sampling)  Connect the pre-charge port to the
DuraSite to the pipeline for pre-charge pressure (Proceed to
step 3d), or configured like the liquid sample application below.
(Step 3c)

3c:  (Light sampling)  Pre-charge the DuraSite as indicated in
Step 1, then install a pressure relief valve to the pre-charge port
and open the pre-charge valve on the DuraSite. (The pressure
relief valve should have a relief pressure setting of
approximately 100 psi above line pressure.)

3d:  Open the product inlet valve of the DuraSite and the purge
valve on the sampler. Next open the purge valve on the product
end of the DuraSite and allow product to purge all lines and
connections out.

3e:  Close purge valves and begin sample cycle.

3f:  At the end of sample cycle, close product inlet valve on the
DuraSite and remove the DuraSite. Pack the DuraSite in
appropriate carrying case to meet D.O.T. guideline, with D.O.T.
paperwork and transport to lab for analysis.

Step 4:  (In The Lab)  Prior to analysis, the product should be
mixed.  This is accomplished simply and efficiently by inverting
the cylinder end-over-end,  causing the mixing ball to fall
through the product.  Approximately 10-12 trips of the mixing ball
through the product assures a homogenous solution.

Step 5:  The regulated pre-charge gas should be reconnected
to the pre-charge side of the cylinder.  The pre-charge gas
supply should remain open during analysis.

Step 6:  Purging a small amount of product from the vessel
removes unmixed product from the tee, relief device, gauge,
etc.  The unit can now be connected to a chromatograph and
the product analyzed.

Step 7:  After analyzing, the remainder of the product should be
dumped and the vessel properly cleaned.  Normal cleaning can
be accomplished by rinsing the product end with a petroleum
solvent and flushing with acetone.  If a more thorough cleaning
is required, the vessel should be disassembled.

WARNING:  A portable sample vessel should never be  filled
to more than 80%.  This allows a 20% pre-charge cushion to
absorb thermal expansion of the product.
Shipping:  Extreme care should be taken when preparing a
vessel for shipment.  Both valves should be capped to prevent
possible leakage.  The vessel should be placed in a
snug-fitting, well-padded and durable case.  All applicable DOT
regulations should be adhered to.

Содержание DynaPak 2010XN

Страница 1: ...DynaPak2010XN SystemSupport Manual Version 08222001...

Страница 2: ...sampling with the DPS 2 12 10 Sampler Set Up proportional to flow sampling 14 11 Sampler Maintenance 16 12 Trouble Shooting 20 13 Diagrams 1 DP 2000 pump assembled 24 2 DP 2000 pump exploded 25 3 YZ...

Страница 3: ...n contact your YZ representative or YZ Customer Service at 936 788 5593 2 System Components The primary components of the DynaPak 2010 Sampling System are illustrated here 1 Introduction Z 65 Controll...

Страница 4: ...l to flow sampling in this mode of operation the Z 65 counter operates as a dividing counter The Z 65 counter periodically energizes a low power solenoid valve As in the other two modes of operation t...

Страница 5: ...could cause aerosols or significant pressure drops 4 2 The sampler should not be located within the defined meter tube region AGA 3 manual A The number of unobstructed straight pipe diameters upstrea...

Страница 6: ...as supply connection for the Filter Regulator located on the left side of the sampler enclosure CAUTION Incorrect operation of valves over tightening can result in damage to the valve components isola...

Страница 7: ...pressure supply valve without ensuring that the equalization valve is open Failure to do so may damage the DPS 2 s internal components d Open the other supply valve e Close the equalization valve f Ru...

Страница 8: ...lume constant pressure cylinder The free floating piston cylinder DuraSite may be installed in a horizontal position on an optional vessel rack Free floating piston cylinders should NOT be installed o...

Страница 9: ...on the top of the pump counterclockwise to set the pump displacement at 4cc stroke 7 4 Move all of the mode switches on the Z 65 to their off positions 7 5 Move both timer counter dials to the 0 posit...

Страница 10: ...E The time 18 minutes above corresponds to the dial setting shown for the Z65 model with the timer range setting in the factory position jumper on the two left pins See section 12 4 Timer Range Settin...

Страница 11: ...8 6 Turn mode switch 3 to on 8 7 Turn mode switch 4 to off 8 8 Press the test button once to initiate the timer sequence 8 3 Adjust the pump volume adjustment knob to the value used in the calculatio...

Страница 12: ...18 minutes above corresponds to the dial setting shown for the Z65 model with the timer range setting in the factory position jumper on the two left pins See section 12 4 Timer Range Setting 3 2 0 9...

Страница 13: ...6 Turn mode switch 3 to off 9 7 Turn mode switch 4 to off 9 8 Press the test button once to initiate the timer sequence NOTE For the Z 65 to operate the DP 2010 System a minimum pressure differential...

Страница 14: ...ng connections to the Z 65 terminal strip see illustration b Turn mode switch 1 to on c Turn mode switch 2 to off d Turn mode switch 3 to on e Turn mode switch 4 to off OR 10 3 If the input is a volta...

Страница 15: ...m ber of pulses you want to count before the sample pump strokes Example 20 pulses turn dials to 20 Press the test button once to load the value into the memory NOTE If the calculated counter setting...

Страница 16: ...ample discharge 1 8 stainless steel tubing from the pump body d Screw the stroke adjustment knob all the way down to the 0 cc stroke setting e Unscrew the pump body by hand from the inlet check valve...

Страница 17: ...ll o rings and bushings to prevent damage j Install the body bushing into the bottom of the pump body NOTE Apply a light coat of assembly grease on the plunger shaft prior to installation k Insert all...

Страница 18: ...must be connected to test the battery condition Battery condition cannot be tested with a volt meter Replacing a Depleted Battery 1 Remove the four thumb screws cover plate and orange terminal connec...

Страница 19: ...for the DynaPak 2000 Series gas samplers Part Number Description Qty Location C4 0004 Filter element 1 see diagrams 3 and 4 D3 0002 DP 2000 pump seal kit 1 see diagrams 1 and 2 D3 0003 YZ filter regu...

Страница 20: ...s sensing flow in the pipeline the green LED should illuminate If flow is not present the red LED should illuminate 12 3 LCD Stroke Indicator Test Mode A To test the stroke counter set the mode switch...

Страница 21: ...factory position far left position 12 4 Timer Range Setting A There are two Z 65 models the Z 65 6 1 and Z 65 6 03 Each Z 65 timer has two ranges for the timer setting dials 1 Z 65 6 1 Range Setting a...

Страница 22: ...minutes set the configuration jumper to the center position Note To obtain maximum battery life choose the longest solenoid stroke rate possible Mutual System Approved Factory factory positioned for 1...

Страница 23: ...green LED will turn on and off and the red LED will illuminate again This will normally occur very quickly and give the appearance that the green LED blinks on when the pulse input is received and re...

Страница 24: ...N B1 0020 Inlet Screen P N C4 0006 Inlet Check Valve Spring Guide Inlet Check Valve Spring O Ring 006 P N A5 1006 O Ring 012 P N A5 1108 O Ring 108 P N B1 0014 Male Connector 1 8 P X 1 8 T P N B1 0015...

Страница 25: ...Housing P N B1 0051 P N C3 0005 P N B1 0030 Set Screw P N C0 0096 P N C0 0014 Cap Screw 6 Each O Ring 012 P N B1 0020 P N C0 0026 Inlet Screen P N C4 0006 P N C3 0008 P N C4 0022 20 Micron Filter Inle...

Страница 26: ...26 Diagram 3 YZ filter regulator assembled...

Страница 27: ...A3 0073 O Ring 012 Retaining Nut P N A5 1012 Seat P N A3 0072 Upper Dart P N A3 0071 Upper Spring P N C3 0010 P N A3 0070 Piston P N A3 0069 Piston Spring Spring Cap P N A3 0068 P N A3 0067 Bonnet Pn...

Страница 28: ...Steel Set Screw Inlet Rupture Disc 1800 psi P N A3 0083 Rupture Disc Retainer P N A3 0084 YZ Valve Body P N A3 0137 0 1000 PSI LinkPlus Model No C1 0005 0 2000 PSI LinkPlus Model No C1 0006 0 600 PSI...

Страница 29: ...0 0005 Set Screw P N E0 5001 Magnet Diaphragm SS Hex Nut P N C0 0048 Qty 8 P N A6 0013 P N G2 0002 Cable Assembly 30 Ft Shrink Wrap SS Bolt P N C0 0047 Qty 8 P N H2 0002 Qty 2 Magnet Support Body Diap...

Страница 30: ...65 6 Controller Assembly 1 1 Model Z 65 6 1 F2 0001 Model Z 65 6 03 F2 0018 2 Battery Pack E3 2001 1 3 Stroke Counter Assembly G1 0001 1 4 Terminal Strip 6 Position H1 0001 1 5 BCD Switch E1 0001 2 6...

Страница 31: ...08222001 ATEX Rev YZ Systems Inc 3101 Pollok Drive Conroe Texas USA 77303 P 936 788 5593 F 936 788 5720 31 Diagram 8 Z 65 Installation Notes Wiring Control Documentation System Installation Notes and...

Страница 32: ...es should be closed The product connection is slowly broken in order to vent any trapped product After vessel removal all connections should be checked for leaks and the pre charge and product valve p...

Страница 33: ...Head YZ UniValve Purge Valve Nut P N A3 0080 P N A3 0063 Valve Seat P N C6 1003 Ser 4608 P N 12 1006 Ser 0321 and Prior P N 12 1007 Ser 0322 0613 P N 12 1008 Ser 0614 4087 P N C6 1002 Ser 4088 Piston...

Страница 34: ...34 Notes...

Страница 35: ...DP 2010XN Version 08222001 ATEX Rev YZ Systems Inc 3101 Pollok Drive Conroe Texas USA 77303 P 936 788 5593 F 936 788 5720 35 Notes...

Страница 36: ...ility and fitness for a particular purpose with respect to the product This limited warranty gives you specific legal rights You may have others which vary from state to state These remedies are not a...

Отзывы: