11
12
y
Average Leak volume
: average air leak volume per minute of the last therapy (L/min).
y
Total Time
: the number of hours of the device working.
y
P90
: the pressure for the 90 percent of the last therapy (cmH
2
O/hPa).
y
Version
: the version of software used for the device.
y
SN
: the serial number of device.
AHI
0.0
Used Time
0.4
Treatment Time
0.4
Average Pressure
20.0
Average Pressure
0.0
Average Leak Volume 0.0
Total Time
37.9
P90
20.0
P90
20.0
Version
V1.00_AIC
SN
YH550A*940100001
Back<<
Picture2-1:Info Picture 2-2:Info Picture 2-3:Info
② Ramp:
Ramp time is designed to make you more comfortable at the beginning of the therapy; you can adjust
the ramp time from 0 to 45 minute with an increment of 5min.
To adjust the Ramp:
y
Highlight the "Ramp" and press dial to see your current setting;
y
Press dial and rotate it to adjust your setting;
y
Press dial to save your change.
Info
Humidity
Setup
Ramp
Ramp
0
Picture 3-1: Home Picture 3-2: Ramp
③ Humidity level:
Humidifier is designed to moisten the air and is designed to make therapy more comfortable. If you
are getting a dry nose or mouth, turn up the humidity. If you are getting any moisture in your mask,
turn down the humidity. You can set Humidity level between 0 to 6, where 0 means this function is
disabled, 1 is the lowest humidity setting and 6 is the highest humidity setting.
The temperature of each level on the heater plate is as following (accuracy is ±4℃ )
0=OFF 1=33℃ 2=35℃ 3=40℃ 4=45℃ 5=50℃ 6=55℃
This temperature has been tested with empty water tank, with 10 minutes preheating time.
To adjust the humidity level:
y
Highlight the "Humidity", press dial to enter the humidity page;
y
Press the dial and rotate it to choose humidity level, then press dial again to save your change and
back to home page;
y
You can change humidity level at any time during therapy.
Info
Humidity
Setup
Ramp
Humidity
6
Picture4-1: Home Picture4-2: Humidty
④ Setting:
y
Sound Reminder
: When this function is enabled, the device will give a reminder, if abnormal air
leakage occurs.
y
Smart Start/Stop
: When this function is enabled, therapy starts automatically when you breathe
into your mask. After you remove your mask, it will stop automatically in one minute.
y
FPS Level
: When this function is enabled, you may feel easier to breathe out; It helps you get used
to the therapy. It has 4 levels of FPS, 0 means this function is disabled, 1 is the lowest level of FPS,