09
10
9
10
9. Rotate the knob cover component from the
to .
10.Connect the air tube firmly to the air outlet
located on the top of the device.
11
12
11.Connect the free end of the air tube firmly
onto the mask; See the mask user guide
for detailed information.
12.Plug the power connector into the rear of
the device.
10. Therapy
f
10.1 Start therapy
① Fit your mask;
② Press
or breathe normally if Smart Start/Stop is enabled, therapy will start. During the
therapy, the real-time average pressure, the set pressure, the ramp time (min), and the humidity
level will all be displayed on the screen;
③ During the ramp time, the pressure increases gradually until the setting pressure has been
reached;
④ The screen backlight will turn off automatically after two minutes of no action. You can press any
button to turn it back on.
Caution:
When the level of the humidity is 1-6 and the device is delivering gas pressure,the heating plate
is heating ( including the situations that the screen shows "Large Leakage" and the backlight of the
screen is off ) . When the level of the humidity is 0, the heat plate is not heating. Also if the device is
not delivering gas pressure, the heat plate will not heat either.
Ramp:
15
Humidity: 6
APAP
6.0-15.0
6.0
Pressure
Picture 1: therapy
f
10.2 Stop therapy
① Remove your mask first;
② Press the
or if Smart Start/Stop is enabled, therapy will stop automatically after about one
minute;
③ To power off your device, please separate your plug from the electricity.
11. Function
f
11.1 Function that patient can safely use
① Information:
You can read summary of sleep report on information page. Following parameters will be displayed
y
AHI
: indicates the number of apneas and hypopneas per hour.
y
Used Time
: the number of hours of the device working last session(h).
y
Treatment Time
: the number of hours of therapy patient received last session(h).
y
Average Pressure
: the average pressure of the last therapy (cmH
2