23
24
16. Specified of technology
Items
Specifies
Power
Powered by adapter (Model: BJE1M-0060-N600/DA-60I24)
Input: 100~240VAC, 50~60Hz, 2A max
Output: 24V DC, 2.5A
Environment
conditions
Temperature
Operate: +5℃ ~35℃(+41
℉
~95
℉
),Non-condensing
Transport: -20℃ ~70℃(-4
℉
~158
℉
)
Storage: -20℃ ~70℃(-4
℉
~158
℉
)
Humidity
Operate: relative humidity 15%~90%, Non-
condensing
Transport: relative humidity 15%~90%
Storage: relative humidity 15%~90%
Atmospheric pressure
range
700hPa~1060hPa
Altitude
≤ 3000m
Protection class
IP 21, Class Ⅱ
Mode of operation
Continuous operation
Maximum single fault
steady pressure
Device will shut down in the presence of a single fault if the steady state
pressure exceeds: 40 cmH
2
O
Sound
Sound pressure level
Sound pressure level measured according to ISO
80601-2-70:2015 (CPAP mode)
BreathCare PAP ≤ 32dB(A)
Sound power level
Sound power level measured according to ISO
80601-2-70:2015 (CPAP mode)
BreathCare PAP ≤ 40dB(A)
Physical properties
Dimensions
(length*width*height)
272.5mm * 181mm * 93mm or 10.73'' * 7.13'' * 3.7''
Weight
About 1419 g
Air tube
Plastic hose,about 1.8m
Maximum volume of
water tank
260±10 mL
Materials of water tank PC, Injection molded plastic, stainless steel and
silicone seal
Air outlet
22 mm (complies with ISO 5356-1:2015)
Temperature
Maximum temperature
of heater plate
55℃ (131
℉
)(±4℃ )
Cut-out
110℃(if it is damaged, return to manufacturer)
Maximum gas
temperature
≤ 41℃
Air filter
Material: Polyester non-woven fiber
Average arrestance: ≥ 85%for~2.5 micron dust
Therapy pressure
4~20 hPa (adjustable, step is 0.5hPa) or 4~20 cmH
2
O
Initial pressure
4~20 hPa (adjustable, step is 0.5hPa) or 4~20 cmH
2
O