33
SYMBOL
MEANING
SYMBOL
MEANING
Type B applied part
Do not expose to rain
Fragile article
Lift up
Attention! Please consult
document attached!
3. SYMBOLS REGARDING SAFETY REQUIREMENT
OF THE MACHINE AND MEANING CONCERNED
4. COMMON FAULT ANALYSIS AND
TROUBLESHOOTING
FAULT
CAUSE ANALYSIS
TROUBLESHOOTING
Too small volume of spray
Medication cup module
is not assembled correctly
Restart after medication cup
module is reassembled
Liquid is used up or does not
contact spray sheet for longer
than 10s
Incline the positive surface of
nebulizer to the user and make
liquid contact spray sheet
Spray sheet is blocked
Clean medication cup accor-
ding to operation instructions;
change a new cup if the fault is
not removed through cleaning
The electrode of main unit or
medication cup is dirty
Restart after cleaning electrode
After power on, the power indi-
cator LED flashes to off
Medication cup module
is not assembled correctly
Restart after medication cup
module is reassembled
Medication cup has no liquid or
liquid is used up
Inject liquid
Liquid does not contact spray
sheet for longer than 10s
Incline the positive surface of
nebulizer to the user and make
liquid contact spray sheet
The electrode of main unit or
medication cup is dirty
Restart after cleaning electrode
EN
Содержание M102
Страница 2: ...2...