30
REMOVE THE BATTERIESAND PULL OUT EXTERNAL POWER ADA-
PTER PRIOR TO CLEANING AND MAINTENANCE.
SINCE MEDICATION CUP MODULE IS CONSUMPTIVE PART, IT IS
BEYOND OUR WARRANTY. IN GENERAL, SERVICE LIFE OF MEDI-
CATION CUP IS HALF A YEAR (THREE TIMES OR 30 MIN EVERY
DAY UNDER NORMAL TEMPERATURE). THE ACTUAL SERVICE LIFE
WILL VARY WITH DRUG TYPE. RESIDUAL LIQUID, IF HARD TO BE
REMOVED, CAN BE CLEANED WITH BOILED HOT WATER OR FOR
MANY TIMES.
CLEAN RESIDUAL LIQUID AFTER USING EVERY DAY; IF NOT,
SPRAY SHEET OF NEBULIZER WILL BE BLOCKED AND NEBULI-
ZATION EFFECT WILL BE AFFECTED. CLEANING MODE IS ONLY
USED IN SPRAY SHEET CLEANING AFTER NEBULIZATION AND
SHOULD NOT BE USED IN NORMAL LIQUID INHALATION
Cleaning
2
•
Main unit: Wipe stains on main
unit gently with wet gauze; then
wipe main unit with new gauze
to make it dry.
•
Clean electrodes on main unit
and medication cup module, so
as to ensure reliable connection
between medication cup module
and main unit.
DO NOT USE VOLATILE LIQUID (E.G. BENZENE, GASOLINE OR
DILUENT) TO WIPE NEBULIZER.
DO NOT TOUCH ELECTRODES ON MAIN UNIT AND MEDICATION
CUP MODULE WITH COTTON SWAB OR OTHER ARTICLES, FOR
FEAR OF ELECTRODE FALLING.
REMOVE BATTERIES PRIOR TO PARTS CLEANING. IF POWER ADA-
PTER IS USED, PULL OFF POWER ADAPTER OFF FROM SOCKET
AFTER POWERING OFF.