Acetic acid solution disinfection:
Disinfect inhalation device with
acetic acid solution with white
vinegar and purified water ratio
for 1:3.
FULLY WASH THE PART WITH PURIFIED WATER AFTER ACETIC
ACID SOLUTION DISINFECTION, SO AS TO AVOID ACETIC ACID
REMAINING.
PREVENT MAIN UNIT AND MEDICATION CUP MODULE FROM FAL-
LING
OR
BEING
STRONGLY
IMPACTED.
DO NOT POKE NEBULIZER WITH PINHEAD OR OTHER SHARP
OBJECT.
DO NOT EXPOSE MAIN UNIT AND OTHER PARTS TO EXTREME
HIGH OR LOW TEMPERATURE OR DIRECT SUNLIGHT.
DO NOT PLUG IN/OUT POWER ADAPTER WITH WET HANDS.
THE PRODUCT IS NOT APPLICABLE TO RESPIRATORY ANESTHE-
SIA SYSTEM OR RESPIRATORY SYSTEM;
DO NOT PLACE NEBULIZER, WHETHER IN DOWNTIME OR OPE-
RATION STATE, IN PLACE WHERE CHILDREN OR MENTALLY ILL
PEOPLE CAN REACH. CHILDREN OR MENTALLY ILL PEOPLE SHO-
ULD NOT USE THE PRODUCT UNLESS SUPERVISED BY ADULT.
NEBULIZER CAN BE USED BY MULTIPLE PERSONS, BUT ACCES-
SORIES (MASK AND MOUTHPIECE) HAVING CONTACT WITH HU-
MAN BODY MUST BE USED ONLY BY ONE PERSON, FOR FEAR OF
CROSS INFECTION. WHERE MULTIPLE PERSONS USE ONE MAIN
UNIT AND MULTIPLE ACCESSORIES, THE ACCESSORIES USED
SHOULD
BE
KEPT
SEPARATELY
WITH
MAIN
UNIT.
DO NOT DISASSEMBLE MAIN UNIT OR TRY TO REPAIR MAIN UNIT.
DO NOT STORE NEBULIZER WITH LIQUID OR WATER IN MEDI-
CATION CUP.