Fig. 3-1: Pressurization  delay setting

Fig. 3-2:Pressurization  delay setting

This function makes the therapy more comfortable. The Ramp refers to the time duration required by 
the pressure to increase from the initial to the minimum gradually or the time duration of the therapy 
pressure. The Ramp range is 0~45min, with an increment of 5min.

How to adjust the Ramp:

 ● Place the cursor on Ramp and press the knob to enter the setting interface.
 ● Turn the knob to select the pressurization delay.
 ● Press the knob again to save the settings and return to the standby interface.

  

There are 0-3 FPS levels; Level 0 is to turn off the function, Level 1 is the lowest level and Level 3 the 
highest. This function can help users to quickly adapt to the therapy of the PAP device.

Fig. 4-1 Humidity setting

Fig. 4-2: Humidity setting

 3. Humidity level

The  heated  humidifier  is  used  to  heat  and  humidify  the  ambient  air,  making  the  therapy  more 
comfortable. During the therapy, if the user feels thirsty, the heated humidifying function can be turned 
on. Reduce the Humidity level if the water vapor is felt inside the mask. There are 0-6 Humidity levels 
with Level 0 for turn off, Level 1 for the lowest level and Level 6 for the highest level. Levels 1-6 are 
equivalent  to  the  following  temperatures  (with  the  accuracy  of  ±4℃)  respectively.  (Note:  These 
temperatures are measured without the humidifier tank working):

Place the cursor on the Humidity option, press the knob to enter into the setting interface.
Turn the knob to select the Humidity level, press the knob to save the settings and return to the stand 
by interface.
Press and turn the knob to adjust the Humidity level, and press the knob again to save the settings 
during the therapy.

06

07

Level 1 = about 33℃,  

Level 2 = about 35℃,  

Level 3 = about 40℃,

Level 4= about 45℃,  

Level 5 = about 50℃,  

Level 6 = about 55℃

 

Humidifier  light:

  Turn  on  the  humidifier  light  in  order  to  see  the  remaining  water  quantity  in  the 

humidifying tank clearly.

Audible alarm: 

When this function is turned on, an audio alarm will be sent incase of an air leak or 

motor failure occurred.

Smart start/stop: 

With this function turned on, the PAP will start automatically when you put on the 

nose mask; and stop within 1 min after you remove the mask. 

Time settings: 

Adjust the time (minute, hour, day, month, year) displayed on the device.

Alarm clock: 

Turn on this function and adjust the alarm time. (

hour, minute

Language: 

English or Chinese can be selected.

5. Other functions

2. RAMP

How to adjust the Humidity level:

4. FPS level

Содержание BreathCare YH-360CPAP

Страница 1: ...certificate or the package for the production date 1030400 0A BreathCare APAP Suzhou Yuyue Medical Technology Co Ltd No 9 Jinfeng Road Suzhou Science Technology Town Suzhou Jiangsu 215163 PRC http ww...

Страница 2: ...sk barotrauma severe lung pneumothorax dehydration cerebrospinal fluid leakage recent intracranial surgery intracranial trauma and etc Dry mouth nosebleeds bloating ear or sinus discomfort eye irritat...

Страница 3: ...wise to decrease START Start Stop button Knob Note Fill the humidifier with water which should not exceed the maximum level MAX otherwise the water may enter into the air hose and the device Please fi...

Страница 4: ...s per hour Used time the time duration for the last therapy h Average pressure the average pressure for the last therapy cmH20 hPa Average leak volume average air leak volume per minute for the last t...

Страница 5: ...n off Level 1 for the lowest level and Level 6 for the highest level Levels 1 6 are equivalent to the following temperatures with the accuracy of 4 respectively Note These temperatures are measured wi...

Страница 6: ...knob to return to the standby interface Fig 6 2 In the APAP mode set the maximum pressure minimum pressure therapy pressure and initial pressure by the same operational order Remark No therapy pressu...

Страница 7: ...return option and press the knob to return to the standby interface Fig 6 2 5 Select the Ramp min 6 Select the Humidity Note Press the START button and the knob at the same to enter into the clinical...

Страница 8: ...ce with a dry and clean cloth Note Ensure that the size of the air filter cotton is proper so as to prevent water or dust from going into the PAP device B Air hose inspection Inspect the air hose for...

Страница 9: ...n the figure The removable cover the partition and the humidifier tank body should be properly installed with their embossing positions aligned 14 15 XII Therapy parameter storage The PAP device shoul...

Страница 10: ...he nose mask type is unqualified The humidify level is low The nose has other disorders The humidify level is too high The temperature difference between inside and outside of the hose is high Decreas...

Страница 11: ...duce the service life of the product When the heated humidifier is used please place the device in a height below the user s head to prevent any water from backflowing into the air hose and the nose m...

Страница 12: ...eated humidifier Plastic hose 1 8m approx PC injection molding stainless steel bottom TPE seal strip 4 20cm H2O hPa adjustable increment 0 5cm H2O hPa 4 20cm H2O hPa adjustable increment 0 5cm H2O hPa...

Страница 13: ...s declaration electromagnetic immunity Immunity test Electrostatic discharge ESD GB T17626 2 6 kV contact 8 kV air 6 kV contact 8 kV air 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycle...

Страница 14: ...MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people...

Страница 15: ...I Warranty Generally Suzhou Yuyue Medical Technology Co Ltd provides the warranty periods for the product and its accessories as follows The warranty for this product applies only for the first time b...

Страница 16: ...ase Date Dealer User Date The following cases are not included in the warranty scope 1 Failure or damage caused by any unauthorized modification and repair by non authorized technicians from yuwell co...

Страница 17: ...nt electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the BreathCare PAP device as recommended below according to t...

Отзывы: