Part name

Material

Please select the disinfection method listed in the following table: 

Sterilization 

by boiling

Ingredient: alcohol

Main disinfectant: 

disinfection with 

ethanol

Mouthpiece

Mask

Nebulizer 
kit

Air tube

Filter

PP

PVC

PP

PVC

PET

×

×

×

×

×

VI. Replacement of filter

VII. Maintenance

Typically,  a  filter  sheet  shall  be  replaced  every  three  months.  If 

the  air  filter  has  changed  color,  replace  it  with  a  new  one.  If 

medication has spilled on the air filter, replace with a new air filter 

immediately.

1.

 

Pull  the  air  filter  cover  to  remove  from  the  back  side  of  the 

compressor.

2. Replace the filter.

Remove  the  old  filter  sheet  with  a  toothpick  or  other  tools  and 

then install a new one.

3. Install the cover of filter sheet.

Caution: 

Before use, make sure the filter sheet is properly installed.

Please use the original filter sheet only and do not use the device 

if the filter sheet is not installed.

Do not attempt to wash or clean the filter sheet. Damp air filters 

may  cause  blockages.  If  water  or  medication  has  spilled  on  the 

filter, replace with a new air filter immediately.

Filter sheet has no positive or negative side.

Please  confirm  whether  the  filter  sheet  is  clean  and  whether 

there is any stained dust before installing.

To prevent clogs, regularly clean the filter cover sheet with water 

(non-boiling) or other methods, dry naturally and install properly.

405A                                                 405B

405A                                                 405B

Содержание 405A

Страница 1: ...al closely JIANGSU YUYUE MEDICAL EQUIPMENT SUPPLY CO LTD Sales Center Address Huanyuan East Road No 1 Xuzhuang Software Park Nanjing Jiangsu Province P R China 210000 Manufacturing Address YunYang Ind...

Страница 2: ...intenance 03 III Sale and others 19 01 Contents 1 Power Input AC100 240V 350mA 50 60Hz Output DC12V 1000mA 2 Maximum nebulization rate 0 2mL min 3 MMD 3 9 m Percentage of nebulized particles with diam...

Страница 3: ...transportation 02 03 1 The pressure source is DC compression pump 2 Innovative designs simple operation compacted size easy storage and portable 3 Schematic diagram of working principles The compresse...

Страница 4: ...or manufacturer timely I Open package Checking II Schematic of each components and names of the whole machine Medication cup Medication cup Power Switch Power Switch Air tube Air tube Power plug Powe...

Страница 5: ...the scale on your syringe or vial for accurate measurement of 2 Add the correct amount of prescribed medication to the medication cup III Preparing the Nebulizer for Use medication After the medicatio...

Страница 6: ...ect the air tube to the connector of the main unit and the medication cup respectively 1 Connect one end to the main unit connector 2 Connect the other end to the medication cup connector So far prepa...

Страница 7: ...zer mask and do not let the medication left in the face Rotating the cap can properly adjust aerosol quantity 3 Hold the nebulizer kit for treatment 1 Cleaning Following cleaning instructions after ea...

Страница 8: ...the disinfectant manufacturer a Disinfectant soaking should be in accordance with the time required in the instruction manual of disinfectant Rinse with clean warm water and allow to air dry in a cle...

Страница 9: ...e filter Remove the old filter sheet with a toothpick or other tools and then install a new one 3 Install the cover of filter sheet Caution Before use make sure the filter sheet is properly installed...

Страница 10: ...is and Troubleshooting Symbol Meaning Type II device Symbol Meaning Type B application No Symptom Analysis 1 The device is abnormally loud The air filter cover is incorrectly attached Power is not on...

Страница 11: ...his description are for proper and safe use of the product and the prevention of harm to the user or others or damages Warnings and notes are in the following For patients in coma or sputum patients p...

Страница 12: ...ipe off the liquid with gauze or other soft absorbent material I Sale 1 If there is any quality problem caused by non human factors within one week from the date of sale the company is responsible for...

Страница 13: ...ltage power supply network The processing of discarded main unit accessories and other products should be in accordance with the provisions of the local government Inform 1 For machine damages and fai...