Caution: 

Make  sure  the  air  tube  is  firmly  connected  to  the  main  unit  and 

the medication cup.

Caution: 

Do  not  pull  the  power  cord  strongly  when  unplugging  it.  Please 

hold the adapter and pull it out.

Caution: 

In  nebulization,  keep  the  mouthpiece  in  mouth  and  close  your 

mouth.

Do not tilt the nebulizer kit so the angle of the kit is greater than 

45 degrees. Medication may flow into the mouth and the spraying 

may be ineffective.

If  there  is  anything  unusual  in  using,  please  stop  using  it 

immediately.

In nebulization, check whether there is visible aerosol discharging 

from  the  medication  cup.  In  the  event  of  irregular  spray,  please 

immediately stop the treatment.

Be sure to wipe the face after using 

nebulizer mask, and do not let the 

medication left in the face.

※ 

Rotating the cap can properly 

adjust aerosol quantity.

3. Hold the nebulizer kit for treatment.

1. Cleaning

Following  cleaning  instructions  after  each  use  will  prevent  any 

remaining  medication  in  the  bottle  from  drying  resulting  in  the 

device nebulizing effectively and will help prevent infections.

First,  remove  the  baffle  and  accessories,  discharge  remaining 

medication  in  medication  cup,  and  do  related  cleaning  or 

disinfection.

1) Baffle, mask and mouthpiece

Rinse with clean water or immerse in warm water for 15 minutes 

4. After treatment, turn the power switch to the off (

) position.

5. Disconnect the nebulizer kit from the air tube.

6. Check the air tube. No condensation or moisture should 

remain in the air tube.

7. Disconnect the air tube.

8. Unplug the adapter from 

the power outlet.

Inhaling with the mouthpiece or mask

405A                                          405B

V

.

Cleaning after Each Use

Содержание 405A

Страница 1: ...al closely JIANGSU YUYUE MEDICAL EQUIPMENT SUPPLY CO LTD Sales Center Address Huanyuan East Road No 1 Xuzhuang Software Park Nanjing Jiangsu Province P R China 210000 Manufacturing Address YunYang Ind...

Страница 2: ...intenance 03 III Sale and others 19 01 Contents 1 Power Input AC100 240V 350mA 50 60Hz Output DC12V 1000mA 2 Maximum nebulization rate 0 2mL min 3 MMD 3 9 m Percentage of nebulized particles with diam...

Страница 3: ...transportation 02 03 1 The pressure source is DC compression pump 2 Innovative designs simple operation compacted size easy storage and portable 3 Schematic diagram of working principles The compresse...

Страница 4: ...or manufacturer timely I Open package Checking II Schematic of each components and names of the whole machine Medication cup Medication cup Power Switch Power Switch Air tube Air tube Power plug Powe...

Страница 5: ...the scale on your syringe or vial for accurate measurement of 2 Add the correct amount of prescribed medication to the medication cup III Preparing the Nebulizer for Use medication After the medicatio...

Страница 6: ...ect the air tube to the connector of the main unit and the medication cup respectively 1 Connect one end to the main unit connector 2 Connect the other end to the medication cup connector So far prepa...

Страница 7: ...zer mask and do not let the medication left in the face Rotating the cap can properly adjust aerosol quantity 3 Hold the nebulizer kit for treatment 1 Cleaning Following cleaning instructions after ea...

Страница 8: ...the disinfectant manufacturer a Disinfectant soaking should be in accordance with the time required in the instruction manual of disinfectant Rinse with clean warm water and allow to air dry in a cle...

Страница 9: ...e filter Remove the old filter sheet with a toothpick or other tools and then install a new one 3 Install the cover of filter sheet Caution Before use make sure the filter sheet is properly installed...

Страница 10: ...is and Troubleshooting Symbol Meaning Type II device Symbol Meaning Type B application No Symptom Analysis 1 The device is abnormally loud The air filter cover is incorrectly attached Power is not on...

Страница 11: ...his description are for proper and safe use of the product and the prevention of harm to the user or others or damages Warnings and notes are in the following For patients in coma or sputum patients p...

Страница 12: ...ipe off the liquid with gauze or other soft absorbent material I Sale 1 If there is any quality problem caused by non human factors within one week from the date of sale the company is responsible for...

Страница 13: ...ltage power supply network The processing of discarded main unit accessories and other products should be in accordance with the provisions of the local government Inform 1 For machine damages and fai...

Отзывы: