Clean and recondition the sensor if a slow response in the field has been reported or if it takes more than 90 
seconds to stabilize in pH buffer. 
 
If you get an error message during a pH calibration, check the following: 
 
1.  Ensure the pH buffers are good and not expired 
2.  Ensure that the pH sensor is installed in the correct port of the cable and the correct ISE is enabled in the 

Sensor Setup menu.   

a.  If using a pH or pH/ORP combo sensor in a 6051020 cable, ensure the sensor is installed in port 1. 
b.  If using a pH or pH/ORP combo sensor in a 60510, 6051020 or 6051030 cable, pH should be 

enabled in ISE1 of the instrument’s Sensor Setup menu. 

c.  If using a pH sensor in a 6051010 or Quatro cable, check to see if the pH sensor is installed in port 1 

or port 2. If the pH sensor is installed in port 1, enable pH in ISE1 of the Sensor Setup menu. If the 
pH sensor is installed in port 2, enable pH in ISE2 of the Sensor Setup menu. Note:  It is not 
recommended to use a pH/ORP combo sensor in 6051010 or Quatro cables. If using a pH/ORP 
combo sensor in a 6051010 or Quatro cable, ORP will not be measured or reported.   

3.  If using a 6051010 or Quatro cable, you must have a sensor installed in port 1 for port 2 to operate.  

Additionally, ensure that the sensor installed in port 1 is in good working order. In 6051010 and Quatro 
cables, the sensors installed in port 1 and port 2 use the reference from the sensor installed in port 1 only.  
Therefore, if the sensor installed in port 1 is not working properly, the readings from the sensor installed in 
port 2 will be erroneous. For greatest ease, install a pH sensor in port 1 of both 6051010 and Quatro cables 
and your other ISE sensor in port 2.   

4.  If you continue to get error messages during calibration, clean and recondition the sensor.   

Cleaning and Reconditioning the pH, ORP or pH/ORP Sensor 

If the pH or pH/ORP sensor has been allowed to dry out or has been stored in distilled or deionized water for an 
extended period of time, soak the sensor in buffer 4 overnight to try and restore functionality.   
 
Cleaning is required whenever deposits or contaminants appear on the glass and/or platinum surfaces or when 
the sensor’s response slows. The cleaning can be chemical and/or mechanical. 
 
Removing the sensor from the cable may make cleaning easier. Initially, moisten a soft clean cloth, lens cleaning 
tissue or cotton swab to remove all foreign material from the glass bulb and/or platinum button. Then use a 
moistened cotton swab to carefully remove any material that may be blocking the reference electrode junction of 
the sensor. 

CAUTION

: When using a cotton swab, be careful NOT to wedge the swab between the guard and 

the glass sensor. If necessary, remove cotton from the swab tip, so that the cotton can reach all parts of the 
sensor tip without stress. You can also use a pipe cleaner for this cleaning if more convenient.  
 
If good pH and/or ORP response is not restored, perform the following additional procedure: 

 

1.  Soak the sensor for 10-15 minutes in clean water containing a few drops of commercial dishwashing liquid. 
2.  GENTLY clean the glass bulb and platinum button by rubbing with a cotton swab soaked in the cleaning 

solution. 

3.  Rinse the sensor in clean water, wipe with a cotton swab moistened with clean water, and then re-rinse with 

clean water. 

 
If good pH and/or ORP response is still not restored, perform the following additional procedure: 
1.  Soak the sensor for 30-60 minutes in one molar (1 M) hydrochloric acid (HCl). This reagent can be 

purchased from most lab supply distributors. Be sure to follow the safety instructions included with the acid. 

Содержание Professional Plus

Страница 1: ...Professional Plus Calibration Tips...

Страница 2: ...on Tips 15 Troubleshooting Tips 16 Nitrate 18 Calibration Tips 18 TroubleShooting Tips 19 Chloride 21 Calibration Tips 21 Troubleshooting Tips 22 Installing and Uninstalling Sensors 24 General Precaut...

Страница 3: ...detailed information on calibration and information on how to setup and operate a Pro Plus please refer to the Pro Plus User Manual Calibration Worksheet The Calibration Worksheet on the following pa...

Страница 4: ...80 from 7 buffer mV value NOTE See pH Cal tips section for additional information Span between pH 4 and 7 and 7 and 10 mV values should be 165 to 180 mV 177 is the ideal distance or 59 mV per pH unit...

Страница 5: ...rush to clean the sensor Quatro Cables Quatro cables have a replaceable combination conductivity temperature sensor p n 005560 All other Pro Plus cables have integral temperature sensors If using a Qu...

Страница 6: ...fferent cable the conductivity sensor is integral to the cable and the sensor has two vent holes located close to the cable Ensure the entire conductivity sensor is submerged in the solution or the in...

Страница 7: ...bathroom foaming tile cleaner with the brush Rinse thoroughly with clean water then check the response and accuracy of the conductivity sensor with calibration solution Quatro Cables Quatro cables hav...

Страница 8: ...uts for each buffer are shown below pH 7 mV value 0 mV 50 mV pH 4 mV value 165 to 180 from 7 buffer mV value pH 10 mV value 165 to 180 from 7 buffer mV value A value of 50 or 50 mVs in buffer 7 does n...

Страница 9: ...to get error messages during calibration clean and recondition the sensor Cleaning and Reconditioning the pH ORP or pH ORP Sensor If the pH or pH ORP sensor has been allowed to dry out or has been sto...

Страница 10: ...m of acid after its disposal and before the disposal of chlorine bleach 1 Soak the sensor for approximately 1 hour in a 1 1 dilution of commercially available chlorine bleach 2 Rinse the sensor with c...

Страница 11: ...erroneous Typical causes for this error message include incorrect Sensor Port setup in the instrument a dirty sensor or bad calibration solution TROUBLESHOOTING TIPS Typical working life for ORP sens...

Страница 12: ...suspect port contamination follow the instructions in the Cleaning a Sensor Port section before reinstalling the sensor 8 If you continue to have problems you can check the offset of the ORP sensor b...

Страница 13: ...not water and that there are not any water droplets on the membrane or temperature sensor 5 After entering the calibration mode wait approximately 5 to 15 minutes for the storage container to become c...

Страница 14: ...is in a blue labeled bottle NaCl 5 Recondition the DO sensor and then install a new membrane 6 If you suspect port contamination remove the sensor and follow the instructions in the Cleaning a Sensor...

Страница 15: ...r deionized water The anode should then be thoroughly wiped with a wet paper towel to remove the residual layer from the anode Trapping residual ammonia under the new membrane cap can quickly tarnish...

Страница 16: ...oxygen even when the Pro Plus is turned off This factor allows the sensor to be used with no warm up time as soon as the instrument is powered on However because the sensor is on all the time some sol...

Страница 17: ...on This may take several hours 4 The ammonium sensor can be calibrated with up to three calibration points For highest accuracy perform a two point calibration with 1 and 100 mg L standards within 10...

Страница 18: ...istilled or deionized water to the flask swirl to dissolve all of the reagents and then dilute to the volumetric mark with distilled or deionized water Mix well by repeated inversion and then transfer...

Страница 19: ...night 6 If you suspect port contamination follow the instructions in the Cleaning a Sensor Port section 7 If you continue to have trouble calibrating the ammonium sensor contact your local YSI Represe...

Страница 20: ...nsor tip and the temperature sensor are submerged in solution If using a Quatro cable or 6051030 cable ensure that the conductivity sensor is also submerged in the calibration solution The salinity re...

Страница 21: ...fate to the flask Add approximately 500 mL of distilled or deionized water swirl to dissolve the solid reagents and then dilute to the volumetric mark with water Mix well by repeated inversion and the...

Страница 22: ...enter Cleaning and Reconditioning the Nitrate Sensor The nitrate sensor uses a PVC membrane As always when handling a sensor care should be taken to avoid damaging the membrane After extensive use the...

Страница 23: ...at the chloride sensor tip and the temperature sensor are submerged in solution If using a Quatro cable or 6051030 cable ensure that the conductivity sensor is also submerged in the calibration soluti...

Страница 24: ...is approximately 3 6 months depending on use storage and maintenance Proper storage and maintenance generally extends the sensor s working life If you get error messages during a chloride calibration...

Страница 25: ...ane This sensor can be regenerated by washing with alcohol and or gently polishing with fine emery paper in a circular motion to remove any deposits or discoloration then thoroughly washing with deion...

Страница 26: ...grease is present remove it using a lens cloth or lint free cloth To remove the o rings Use a small flat bladed screwdriver or similar blunt tipped tool to remove the o ring from its groove Do not use...

Страница 27: ...es in the stainless steel retaining nut Next use the installation tool to turn the stainless steel retaining nut counter clockwise to loosen Do not allow the sensor to be turned with the tool Turning...

Страница 28: ...the Pro Plus display to the value of the solution If the readings have drifted more than the accuracy specification of the sensor perform a calibration before taking field measurements YSI Confidence...

Отзывы: