Installing and Uninstalling Sensors 

GENERAL PRECAUTIONS 

It is important that the entire sensor connector and cable connector be dry when installing, removing or replacing 
sensors. This will prevent water from entering the port. Once a sensor is removed, examine the connector inside 
the port. If any moisture is present, use compressed air to completely dry the connector or place directly in front 
of a steady flow of fresh air. If you suspect port contamination, follow the port cleaning procedures listed under 
Cleaning a Sensor Port.   
 
Remove sensors upside down (facing the ground) to help prevent water from entering the port upon removal. 
 
The instrument utilizes o-rings as seals to prevent water from entering the sensor ports. When the sensors are 
removed, the o-rings that provide the seal should be carefully inspected for contamination (e.g. debris, grit, etc.) 
and cleaned if necessary.   
 
If no dirt or damage to the o-rings is evident, wipe the o-rings with a lint free cloth or lens cloth to remove the old 
o-ring grease. Then, lightly apply new o-ring grease (provided in the maintenance kit) to the o-rings without 
removing them from their groove. If there is any indication of damage, the o-ring should be replaced with an 
identical o-ring. At the time of o-ring replacement, the entire o-ring assembly should be cleaned. 
 
Do not over-grease the o-rings. The purpose of the o-ring grease it to keep the o-ring in good condition.  Excess 
grease may collect grit particles that can compromise the seal. Excess grease can also cause the waterproofing 
capabilities of the o-ring to diminish, potentially causing leaks. If excess grease is present, remove it using a lens 
cloth or lint-free cloth.  
 

To remove the o-rings: 

Use a small, flat-bladed screwdriver or similar blunt-tipped tool to remove the o-ring from its groove. Do not use a 
sharp object to remove the o-rings. Using a sharp object could damage the o-ring groove which would allow 
water to enter the port resulting in permanent damage to the port and sensor. Check the o-ring and the groove 
for any excess grease or contamination. If contamination is evident, clean the o-ring and nearby plastic parts 
with lens cleaning tissue or equivalent lint-free cloth. Alcohol can be used to clean the plastic parts, but use only 
water and mild detergent on the o-ring itself.  Using alcohol on o-rings may cause a loss of elasticity and may 
promote cracking. Also, inspect the o-rings for nicks and imperfections. 
 
Before re-installing the o-rings, make sure to use a clean workspace, clean hands, and avoid contact with 
anything that may leave fibers on the o-ring or grooves. Even a very small amount of contamination (hair, grit, 
etc.) may cause a leak. 
  

To re-install the o-rings: 

Place a small amount of o-ring grease between your thumb and index finger. Draw the o-ring through the grease 
while pressing the fingers together to place a very light covering of grease to the o-ring. Place the o-ring into its 
groove making sure that it does not twist or roll.  Do no excessively stretch the o-ring during installation.   
  
Use your grease-coated finger to once again lightly go over the mating surface of the o-ring.   
 
Do not over-grease the o-rings. The excess grease may collect grit particles that can compromise the seal. 
Excess grease can also cause the waterproofing capabilities of the o-ring to diminish, potentially causing leaks. If 
excess grease is present, remove it using a lens cloth or lint-free cloth.  
 
 

24 

Содержание Professional Plus

Страница 1: ...Professional Plus Calibration Tips...

Страница 2: ...on Tips 15 Troubleshooting Tips 16 Nitrate 18 Calibration Tips 18 TroubleShooting Tips 19 Chloride 21 Calibration Tips 21 Troubleshooting Tips 22 Installing and Uninstalling Sensors 24 General Precaut...

Страница 3: ...detailed information on calibration and information on how to setup and operate a Pro Plus please refer to the Pro Plus User Manual Calibration Worksheet The Calibration Worksheet on the following pa...

Страница 4: ...80 from 7 buffer mV value NOTE See pH Cal tips section for additional information Span between pH 4 and 7 and 7 and 10 mV values should be 165 to 180 mV 177 is the ideal distance or 59 mV per pH unit...

Страница 5: ...rush to clean the sensor Quatro Cables Quatro cables have a replaceable combination conductivity temperature sensor p n 005560 All other Pro Plus cables have integral temperature sensors If using a Qu...

Страница 6: ...fferent cable the conductivity sensor is integral to the cable and the sensor has two vent holes located close to the cable Ensure the entire conductivity sensor is submerged in the solution or the in...

Страница 7: ...bathroom foaming tile cleaner with the brush Rinse thoroughly with clean water then check the response and accuracy of the conductivity sensor with calibration solution Quatro Cables Quatro cables hav...

Страница 8: ...uts for each buffer are shown below pH 7 mV value 0 mV 50 mV pH 4 mV value 165 to 180 from 7 buffer mV value pH 10 mV value 165 to 180 from 7 buffer mV value A value of 50 or 50 mVs in buffer 7 does n...

Страница 9: ...to get error messages during calibration clean and recondition the sensor Cleaning and Reconditioning the pH ORP or pH ORP Sensor If the pH or pH ORP sensor has been allowed to dry out or has been sto...

Страница 10: ...m of acid after its disposal and before the disposal of chlorine bleach 1 Soak the sensor for approximately 1 hour in a 1 1 dilution of commercially available chlorine bleach 2 Rinse the sensor with c...

Страница 11: ...erroneous Typical causes for this error message include incorrect Sensor Port setup in the instrument a dirty sensor or bad calibration solution TROUBLESHOOTING TIPS Typical working life for ORP sens...

Страница 12: ...suspect port contamination follow the instructions in the Cleaning a Sensor Port section before reinstalling the sensor 8 If you continue to have problems you can check the offset of the ORP sensor b...

Страница 13: ...not water and that there are not any water droplets on the membrane or temperature sensor 5 After entering the calibration mode wait approximately 5 to 15 minutes for the storage container to become c...

Страница 14: ...is in a blue labeled bottle NaCl 5 Recondition the DO sensor and then install a new membrane 6 If you suspect port contamination remove the sensor and follow the instructions in the Cleaning a Sensor...

Страница 15: ...r deionized water The anode should then be thoroughly wiped with a wet paper towel to remove the residual layer from the anode Trapping residual ammonia under the new membrane cap can quickly tarnish...

Страница 16: ...oxygen even when the Pro Plus is turned off This factor allows the sensor to be used with no warm up time as soon as the instrument is powered on However because the sensor is on all the time some sol...

Страница 17: ...on This may take several hours 4 The ammonium sensor can be calibrated with up to three calibration points For highest accuracy perform a two point calibration with 1 and 100 mg L standards within 10...

Страница 18: ...istilled or deionized water to the flask swirl to dissolve all of the reagents and then dilute to the volumetric mark with distilled or deionized water Mix well by repeated inversion and then transfer...

Страница 19: ...night 6 If you suspect port contamination follow the instructions in the Cleaning a Sensor Port section 7 If you continue to have trouble calibrating the ammonium sensor contact your local YSI Represe...

Страница 20: ...nsor tip and the temperature sensor are submerged in solution If using a Quatro cable or 6051030 cable ensure that the conductivity sensor is also submerged in the calibration solution The salinity re...

Страница 21: ...fate to the flask Add approximately 500 mL of distilled or deionized water swirl to dissolve the solid reagents and then dilute to the volumetric mark with water Mix well by repeated inversion and the...

Страница 22: ...enter Cleaning and Reconditioning the Nitrate Sensor The nitrate sensor uses a PVC membrane As always when handling a sensor care should be taken to avoid damaging the membrane After extensive use the...

Страница 23: ...at the chloride sensor tip and the temperature sensor are submerged in solution If using a Quatro cable or 6051030 cable ensure that the conductivity sensor is also submerged in the calibration soluti...

Страница 24: ...is approximately 3 6 months depending on use storage and maintenance Proper storage and maintenance generally extends the sensor s working life If you get error messages during a chloride calibration...

Страница 25: ...ane This sensor can be regenerated by washing with alcohol and or gently polishing with fine emery paper in a circular motion to remove any deposits or discoloration then thoroughly washing with deion...

Страница 26: ...grease is present remove it using a lens cloth or lint free cloth To remove the o rings Use a small flat bladed screwdriver or similar blunt tipped tool to remove the o ring from its groove Do not use...

Страница 27: ...es in the stainless steel retaining nut Next use the installation tool to turn the stainless steel retaining nut counter clockwise to loosen Do not allow the sensor to be turned with the tool Turning...

Страница 28: ...the Pro Plus display to the value of the solution If the readings have drifted more than the accuracy specification of the sensor perform a calibration before taking field measurements YSI Confidence...

Отзывы: