YSI Professional Plus Скачать руководство пользователя страница 13

Dissolved Oxygen 

The dissolved oxygen sensor should be calibrated every day the instrument is used. It is not necessary to 
calibrate in both % and mg/L or ppm. Calibrating in % will simultaneously calibrate mg/L and ppm and vice versa.   

CALIBRATION TIPS  

1.  The Pro Plus can be calibrated in air-saturated water, water-saturated air or against a Winkler Titration. You 

can perform a 1 or 2 point DO calibration. A 2 point calibration includes 1 point in a zero oxygen environment 
and the 2

nd

 point at full saturation.   

2.  For both ease of use and accuracy, YSI recommends that you perform a 1 point calibration in water-

saturated air.   

3.  Make sure that there is a good membrane with fresh electrolyte (O2 probe solution) installed on the DO 

sensor. The membrane should be clean and free of wrinkles. There should not be any air bubbles present 
under the membrane. Membranes should be changed regularly and generally last 2-8 weeks depending on 
use and storage.   

4.  To perform a 1 point calibration in water-saturated air, place the sensor in a 100% humid environment. This 

can be accomplished several ways: 

a.  For the 60520 and 6052030 cables, moisten the sponge in the gray calibration sleeve with a 

small

 

amount of clean water and place it over the sensor guard.   

b.  For the 6051020 and Quatro cables, place a small amount of water in the calibration/storage cup and 

place it over the sensors. When screwing the calibration cup onto the sensor bulkhead, only engage 
one or two threads. Do not screw the calibration cup completely onto the sensor bulkhead. The goal 
is to have air exchange between inside and outside the calibration cup.   

The sponge and calibration sleeve/cup should be clean since bacterial growth may consume oxygen and 
interfere with the calibration. Be sure the sensor is in air, not water, and that there are not any water droplets 
on the membrane or temperature sensor.   

5.  After entering the % calibration mode, wait approximately 5 to 15 minutes for the storage container to 

become completely saturated and, if using a polarographic sensor, to allow the sensor to stabilize. 

6.  Salinity affects the ability of water to hold oxygen and is used by the instrument to calculate DO mg/L (ppm). 

The Salinity value displayed near the top of the DO calibration screen is either the salinity correction value 
entered in the Sensor menu or the Salinity value as measured by the conductivity sensor in use. If you are 
using a conductivity sensor, ensure that it is calibrated and reading correctly in order to obtain accurate DO 
mg/L (ppm) measurements. If you are not using a conductivity sensor, the Salinity correction value should be 
the salinity of the water you will be testing. Highlight Salinity and press enter to modify this setting if 
necessary. The salinity of fresh water is typically 0-0.5 ppt and seawater is typically 35 ppt. 

7.  After accepting the calibration, navigate to the GLP menu and record the DO sensor’s value (sensor current 

in uA). The acceptable sensor currents when calibration is performed at 25°C, in a 100% saturated air 
environment at 760 mmHg are: 

 

1.25 mil PE membrane (yellow membrane): Average 6.15 uA (min. 4.31 uA, max. 8.00 uA) 

 

2.0 mil PE membrane (blue membrane): Average 3.38 uA (min. 2.37 uA, max. 4.40 uA) 

 

1 mil Teflon membrane: Average 16.29 uA (min. 11.40 uA, max. 21.18 uA) 

8.  If you receive a warning message stating that the calibration is questionable, do not continue with the 

calibration. Instead, select ‘No’ and investigate what is causing the questionable results. If you accept a 
questionable calibration, your DO readings will be erroneous. Typical causes of a calibration error message 
include:  incorrect sensor, membrane or port setup in the instrument, incorrect barometric pressure 
information, a bad membrane or a sensor that needs reconditioned.   

 

11 

Содержание Professional Plus

Страница 1: ...Professional Plus Calibration Tips...

Страница 2: ...on Tips 15 Troubleshooting Tips 16 Nitrate 18 Calibration Tips 18 TroubleShooting Tips 19 Chloride 21 Calibration Tips 21 Troubleshooting Tips 22 Installing and Uninstalling Sensors 24 General Precaut...

Страница 3: ...detailed information on calibration and information on how to setup and operate a Pro Plus please refer to the Pro Plus User Manual Calibration Worksheet The Calibration Worksheet on the following pa...

Страница 4: ...80 from 7 buffer mV value NOTE See pH Cal tips section for additional information Span between pH 4 and 7 and 7 and 10 mV values should be 165 to 180 mV 177 is the ideal distance or 59 mV per pH unit...

Страница 5: ...rush to clean the sensor Quatro Cables Quatro cables have a replaceable combination conductivity temperature sensor p n 005560 All other Pro Plus cables have integral temperature sensors If using a Qu...

Страница 6: ...fferent cable the conductivity sensor is integral to the cable and the sensor has two vent holes located close to the cable Ensure the entire conductivity sensor is submerged in the solution or the in...

Страница 7: ...bathroom foaming tile cleaner with the brush Rinse thoroughly with clean water then check the response and accuracy of the conductivity sensor with calibration solution Quatro Cables Quatro cables hav...

Страница 8: ...uts for each buffer are shown below pH 7 mV value 0 mV 50 mV pH 4 mV value 165 to 180 from 7 buffer mV value pH 10 mV value 165 to 180 from 7 buffer mV value A value of 50 or 50 mVs in buffer 7 does n...

Страница 9: ...to get error messages during calibration clean and recondition the sensor Cleaning and Reconditioning the pH ORP or pH ORP Sensor If the pH or pH ORP sensor has been allowed to dry out or has been sto...

Страница 10: ...m of acid after its disposal and before the disposal of chlorine bleach 1 Soak the sensor for approximately 1 hour in a 1 1 dilution of commercially available chlorine bleach 2 Rinse the sensor with c...

Страница 11: ...erroneous Typical causes for this error message include incorrect Sensor Port setup in the instrument a dirty sensor or bad calibration solution TROUBLESHOOTING TIPS Typical working life for ORP sens...

Страница 12: ...suspect port contamination follow the instructions in the Cleaning a Sensor Port section before reinstalling the sensor 8 If you continue to have problems you can check the offset of the ORP sensor b...

Страница 13: ...not water and that there are not any water droplets on the membrane or temperature sensor 5 After entering the calibration mode wait approximately 5 to 15 minutes for the storage container to become c...

Страница 14: ...is in a blue labeled bottle NaCl 5 Recondition the DO sensor and then install a new membrane 6 If you suspect port contamination remove the sensor and follow the instructions in the Cleaning a Sensor...

Страница 15: ...r deionized water The anode should then be thoroughly wiped with a wet paper towel to remove the residual layer from the anode Trapping residual ammonia under the new membrane cap can quickly tarnish...

Страница 16: ...oxygen even when the Pro Plus is turned off This factor allows the sensor to be used with no warm up time as soon as the instrument is powered on However because the sensor is on all the time some sol...

Страница 17: ...on This may take several hours 4 The ammonium sensor can be calibrated with up to three calibration points For highest accuracy perform a two point calibration with 1 and 100 mg L standards within 10...

Страница 18: ...istilled or deionized water to the flask swirl to dissolve all of the reagents and then dilute to the volumetric mark with distilled or deionized water Mix well by repeated inversion and then transfer...

Страница 19: ...night 6 If you suspect port contamination follow the instructions in the Cleaning a Sensor Port section 7 If you continue to have trouble calibrating the ammonium sensor contact your local YSI Represe...

Страница 20: ...nsor tip and the temperature sensor are submerged in solution If using a Quatro cable or 6051030 cable ensure that the conductivity sensor is also submerged in the calibration solution The salinity re...

Страница 21: ...fate to the flask Add approximately 500 mL of distilled or deionized water swirl to dissolve the solid reagents and then dilute to the volumetric mark with water Mix well by repeated inversion and the...

Страница 22: ...enter Cleaning and Reconditioning the Nitrate Sensor The nitrate sensor uses a PVC membrane As always when handling a sensor care should be taken to avoid damaging the membrane After extensive use the...

Страница 23: ...at the chloride sensor tip and the temperature sensor are submerged in solution If using a Quatro cable or 6051030 cable ensure that the conductivity sensor is also submerged in the calibration soluti...

Страница 24: ...is approximately 3 6 months depending on use storage and maintenance Proper storage and maintenance generally extends the sensor s working life If you get error messages during a chloride calibration...

Страница 25: ...ane This sensor can be regenerated by washing with alcohol and or gently polishing with fine emery paper in a circular motion to remove any deposits or discoloration then thoroughly washing with deion...

Страница 26: ...grease is present remove it using a lens cloth or lint free cloth To remove the o rings Use a small flat bladed screwdriver or similar blunt tipped tool to remove the o ring from its groove Do not use...

Страница 27: ...es in the stainless steel retaining nut Next use the installation tool to turn the stainless steel retaining nut counter clockwise to loosen Do not allow the sensor to be turned with the tool Turning...

Страница 28: ...the Pro Plus display to the value of the solution If the readings have drifted more than the accuracy specification of the sensor perform a calibration before taking field measurements YSI Confidence...

Отзывы: