YSI ProDSS Скачать руководство пользователя страница 7

YSI ProDSS Meter Calibration and Maintenance 

K:\Streamkeepers\Monitoring\Eqpt & supplies\Calibration-Maintenance\SOPs\YSI ProDSS maint-
cal SOP's  
 

7

Barometer Calibration

 

 
The YSI ProDSS has a built in barometer that has been factory calibrated and rarely needs to be 
recalibrated.  The accuracy of the barometer is essential for dissolved oxygen calibration.  Before 
and after each season, check the barometer readings for accuracy by testing against the weather 
station located in the Roads Department upstairs. Record on Barometer data sheet the weather 
station location, the weather station barometric pressure reading to 0.01 inches Hg, and the 
ProDSS barometer reading to 0.01 inches Hg. 
If the ProDSS’s barometer does not match the value of the weather station it will need 
recalibration before QC-verified data can be collected in the field. 
 

Calibration:

 

1.

 

If barometer units are not in inHg, refer to 

Check Sensor Readings

 on page 4 to change units. 

2.

 

Select the 

calibration button

, and select Barometer. 

3.

 

Select Calibration Value then enter the correct “true” barometric pressure. 

4.

 

Select Accept Calibration 

 
Barometric Unit Conversions: 
• BP in mmHg=25.4 x BP inHg 
• BP in mmHg=0.750062 x BP mb 
• BP in mmHg=51.7149 x BP psi 
• BP in mmHg=7.50062 x BP kPa 
• BP in mmHg=760 x BP atm 

 
 

Conductivity Calibration 

 

With this conductivity/temp sensor, once specific conductance is calibrated, all other parameters 
measured by this sensor will be calibrated.   

 
Equipment Needed: 

 

“Conductivity Calibration Protocol & Data Sheets” notebook  

 

YSI ProDSS meter 

 

Purified water 

 

Mild solution of dish soap and tap water 

 

Cleaning brush found in maintenance kit 

 

Conductivity standard solution 

*Check date if opened. The solution expires one month 

after opening. If expired, get new solution. NEVER REUSE CALIBRATION 
SOLUTION.

 

 

 
 
 
 
 

Содержание ProDSS

Страница 1: ...Sensor Installation Sensor installation Calibration Barometer Conductivity pH Turbidity Dissolved oxygen Maintenance Short term storage Long term storage A FEW NOTES BEFORE GETTING STARTED The instru...

Страница 2: ...and unit short term sensors are still attached to the unit and long term sensors have been removed and individually stored If the unit has been stored under the short term procedures continue to the C...

Страница 3: ...y tighten the retaining nut if any resistance is felt loosen the nut and try again Finger tighten as much as you can then use the sensor installation tool to tighten until snug see image below DO NOT...

Страница 4: ...d Once aligned push the sensor into the bulkhead until it is firmly seated in the port see image below 5 Carefully tighten the retaining nut if any resistance is felt loosen the nut and try again Fing...

Страница 5: ...o finish charging When using the AC adapter it takes approximately 14 hours to charge the ProDSS battery when the instrument is turned off during the charge The amount of time required to completely c...

Страница 6: ...b practices This includes what was calibrated who calibrated it and when it was calibrated To prevent contamination always calibrate the sensors in the following order Barometer if necessary conductiv...

Страница 7: ...fore QC verified data can be collected in the field Calibration 1 If barometer units are not in inHg refer to Check Sensor Readings on page 4 to change units 2 Select the calibration button and select...

Страница 8: ...it sit for about 30 seconds Use the cleaning brush from the maintenance kit to scrub the inside of the chamber insert in holes and scrub back and forth about 20 30 times Rinse the probe in clean tap...

Страница 9: ...sor pH Calibration Equipment Needed pH Calibration Protocol Data Sheets notebook YSI ProDSS meter Purified water Kimwipes pH 4 7 and 10 buffer solutions A note about pH millivolts mV Millivolts mV are...

Страница 10: ...ed meter Record results on the data sheet End of Season pH Calibration Check 1 Turn on and warm up instrument 2 Check and record readings of pH 7 and temp for the instrument in its kit pH 7 buffer Do...

Страница 11: ...oint 2 you are now ready to calibrate with the next buffer solution DO NOT back out to the Run screen by hitting the left arrow button or the Esc button If you don t finish the three point calibration...

Страница 12: ...ffer temperature reading value may be located on pH buffer bottle 7 Confirm stabilization by watching the actual readings on the screen if pH remains within 0 01 for 2 minutes the sensor is stable 8 A...

Страница 13: ...e sensors and calibration cup with zero standard or deionized water Shake off the unit and cup after rinse 3 SLOWLY pour the deionized water down the side of the cup like you re pouring a beer fill th...

Страница 14: ...in that range must be flagged an estimate Q in Streamkeepers flagging system and J or one of its variants in the WA Dept of Ecology s flagging system And if the meter does not perform to 1 0 or 10 of...

Страница 15: ...d on the screen 8 Record results on the data sheet 9 Rinse the probe thoroughly with DI water before placing it back in the calibration cup this is important or it could mess up future calibrations 10...

Страница 16: ...is essential to acceptance of our data by the State Equipment needed 1 or 2 large plastic bucket s depends on method used below DI water from lab DO Barometer Calibration Protocol Data Sheets notebook...

Страница 17: ...the samples were taken on the data sheet WINKLER TITRATIONS IN THE LABORATORY Equipment for completing Winkler titrations Graduated cylinder burette with stopcock at bottom 25 mL with 3 way stopcock...

Страница 18: ...y away from you toward the wall 12 Tighten the valve on the rubber bulb and pressurize the reservoir Pump the rubber bulb gently to raise the solution drain refill and drain twice more to rinse then f...

Страница 19: ...tes and amounts of all solutions used to ensure there s enough unexpired liquids for the next calibration activity If not inform the office staff immediately Meter Samples 1 Place the bucket of prepar...

Страница 20: ...on the ProDSS and its sensors follow the specified procedures in this section which have been taken directly out of the manual or it could void the warranty For more information on the warranty consu...

Страница 21: ...f force 1 Remove the sensor from the bulkhead and soak the end of the sensor in a mild solution of clean tap water and dish soap in a glass beaker Soak for 10 to 15 minutes 2 After soaking rinse with...

Страница 22: ...for installation Replacing the DO Cap NOTE Use extreme care to not damage the threads at the end of the sensor 1 Turn the old cap on the sensor counterclockwise and remove do not use any hand tools to...

Страница 23: ...ly for correct transcription from the Calibration Code Label prior to selecting Update Sensor Cap Coefficients If you receive an error message after several entry attempts contact YSI Technical Suppor...

Страница 24: ...ored more securely and under their own individual proper storing conditions Storage caps can be found in the individual storage boxes of each sensor do not lose these caps The following are a few guid...

Страница 25: ...s only 4 Place the red cap onto the sensor connector end 5 Take out the storage bottle open lid and o ring from the storage box 6 Carefully slide the open lid with the o ring placed inside the lid ont...

Отзывы: