YSI ProDSS Meter Calibration and Maintenance 

K:\Streamkeepers\Monitoring\Eqpt & supplies\Calibration-Maintenance\SOPs\YSI ProDSS maint-
cal SOP's  
 

5

Installing the Sensor Guard

 

1.  Carefully slide the sensor guard over the bulkhead and attached sensors/port 

plugs. Push the sensor guard toward the bulkhead until the sensor guard 
threads align with the bulkhead threads (see image right). 

2.  Carefully finger-tighten the sensor guard clockwise. 

NOTICE:

 

If any resistance is felt, loosen the sensor guard completely to prevent 

cross-threading. Incorrect installation may cause damage to the sensor guard or 
bulkhead that is not covered by the warranty. 
 

Battery Life

 

Use YSI designated charging equipment only! 
If the battery is not charged, the meter must be charged from the AC power 
adapter (strongest charge source), directly from a computer USB connection or 
from an external, portable USB battery pack.  For the instrument to recognize 
that it is using AC power, you must start charging the ProDSS while on. After 
the instrument recognizes it is being charged, it can be turned off to finish 
charging.  When using the AC adapter, it takes approximately 14 hours to charge 
the ProDSS battery when the instrument is turned off during the charge. The 
amount of time required to completely charge the battery pack when the ProDSS 
is initially turned on during the charge is approximately 9 hours.  Expected battery life: 
• ProDSS instrument only - 48 hours 
• ProDSS with fully loaded cable assembly and      25% (Default) LCD brightness - 20 hours 
• ProDSS with fully loaded cable assembly and 100% LCD brightness - 14 hours                   
For maximum battery life, keep the battery 40% to 80% charged. Also, a larger 
discharge (e.g. to 50%) is better than a small discharge (e.g. to 90%) between recharges.   

 
Check Sensor Readings 

After installing all the sensors turn the unit on.  In the 
“Run” screen (main display screen, image right) make 
sure all parameters are being displayed and are 
displaying the proper units: 
temperature - °C 
barometer – inHg 
dissolved oxygen – DO%  
dissolved oxygen – DO mg/L 
conductivity – SPC-µS/cm 
salinity – SAL-ppt 
turbidity – FNU 
If any of these are not right, do the following: 
 

Select the 

probe button 

and scroll down to Display 

submenu and hit enter.  Scroll down and select the 
desired parameter.  A submenu will appear with 
multiple unit options, select the proper unit and return 
to the Run Screen. 
 

 

Содержание ProDSS

Страница 1: ...Sensor Installation Sensor installation Calibration Barometer Conductivity pH Turbidity Dissolved oxygen Maintenance Short term storage Long term storage A FEW NOTES BEFORE GETTING STARTED The instru...

Страница 2: ...and unit short term sensors are still attached to the unit and long term sensors have been removed and individually stored If the unit has been stored under the short term procedures continue to the C...

Страница 3: ...y tighten the retaining nut if any resistance is felt loosen the nut and try again Finger tighten as much as you can then use the sensor installation tool to tighten until snug see image below DO NOT...

Страница 4: ...d Once aligned push the sensor into the bulkhead until it is firmly seated in the port see image below 5 Carefully tighten the retaining nut if any resistance is felt loosen the nut and try again Fing...

Страница 5: ...o finish charging When using the AC adapter it takes approximately 14 hours to charge the ProDSS battery when the instrument is turned off during the charge The amount of time required to completely c...

Страница 6: ...b practices This includes what was calibrated who calibrated it and when it was calibrated To prevent contamination always calibrate the sensors in the following order Barometer if necessary conductiv...

Страница 7: ...fore QC verified data can be collected in the field Calibration 1 If barometer units are not in inHg refer to Check Sensor Readings on page 4 to change units 2 Select the calibration button and select...

Страница 8: ...it sit for about 30 seconds Use the cleaning brush from the maintenance kit to scrub the inside of the chamber insert in holes and scrub back and forth about 20 30 times Rinse the probe in clean tap...

Страница 9: ...sor pH Calibration Equipment Needed pH Calibration Protocol Data Sheets notebook YSI ProDSS meter Purified water Kimwipes pH 4 7 and 10 buffer solutions A note about pH millivolts mV Millivolts mV are...

Страница 10: ...ed meter Record results on the data sheet End of Season pH Calibration Check 1 Turn on and warm up instrument 2 Check and record readings of pH 7 and temp for the instrument in its kit pH 7 buffer Do...

Страница 11: ...oint 2 you are now ready to calibrate with the next buffer solution DO NOT back out to the Run screen by hitting the left arrow button or the Esc button If you don t finish the three point calibration...

Страница 12: ...ffer temperature reading value may be located on pH buffer bottle 7 Confirm stabilization by watching the actual readings on the screen if pH remains within 0 01 for 2 minutes the sensor is stable 8 A...

Страница 13: ...e sensors and calibration cup with zero standard or deionized water Shake off the unit and cup after rinse 3 SLOWLY pour the deionized water down the side of the cup like you re pouring a beer fill th...

Страница 14: ...in that range must be flagged an estimate Q in Streamkeepers flagging system and J or one of its variants in the WA Dept of Ecology s flagging system And if the meter does not perform to 1 0 or 10 of...

Страница 15: ...d on the screen 8 Record results on the data sheet 9 Rinse the probe thoroughly with DI water before placing it back in the calibration cup this is important or it could mess up future calibrations 10...

Страница 16: ...is essential to acceptance of our data by the State Equipment needed 1 or 2 large plastic bucket s depends on method used below DI water from lab DO Barometer Calibration Protocol Data Sheets notebook...

Страница 17: ...the samples were taken on the data sheet WINKLER TITRATIONS IN THE LABORATORY Equipment for completing Winkler titrations Graduated cylinder burette with stopcock at bottom 25 mL with 3 way stopcock...

Страница 18: ...y away from you toward the wall 12 Tighten the valve on the rubber bulb and pressurize the reservoir Pump the rubber bulb gently to raise the solution drain refill and drain twice more to rinse then f...

Страница 19: ...tes and amounts of all solutions used to ensure there s enough unexpired liquids for the next calibration activity If not inform the office staff immediately Meter Samples 1 Place the bucket of prepar...

Страница 20: ...on the ProDSS and its sensors follow the specified procedures in this section which have been taken directly out of the manual or it could void the warranty For more information on the warranty consu...

Страница 21: ...f force 1 Remove the sensor from the bulkhead and soak the end of the sensor in a mild solution of clean tap water and dish soap in a glass beaker Soak for 10 to 15 minutes 2 After soaking rinse with...

Страница 22: ...for installation Replacing the DO Cap NOTE Use extreme care to not damage the threads at the end of the sensor 1 Turn the old cap on the sensor counterclockwise and remove do not use any hand tools to...

Страница 23: ...ly for correct transcription from the Calibration Code Label prior to selecting Update Sensor Cap Coefficients If you receive an error message after several entry attempts contact YSI Technical Suppor...

Страница 24: ...ored more securely and under their own individual proper storing conditions Storage caps can be found in the individual storage boxes of each sensor do not lose these caps The following are a few guid...

Страница 25: ...s only 4 Place the red cap onto the sensor connector end 5 Take out the storage bottle open lid and o ring from the storage box 6 Carefully slide the open lid with the o ring placed inside the lid ont...

Отзывы: