Symptom

Possible Solution

Temperature values 
display Over or Undr 
on Run screen.

1.  Sample temperature is less than -5° C or more 

than +55°C. Increase or decrease the sample 
temperature to bring within the allowable 
range.

2.  Contact YSI Tech Support. Possible temperature 

sensor failure.

Instrument will not 
calibrate pH or ORP; 
instrument displays 
“Calibration Over”, 
“Calibration Under”, 
or “Unstable Reading” 
during calibration.

1.  Verify correct ISE sensor type selection 

in the System Setup menu.  

2.  Verify the calibration solution is accurate.
3.  If calibrating pH, make sure you 

are calibrating buffer 7 first. 

4.  Clean the pH or ORP sensor.
5.  Contact YSI Tech Support.

pH or ORP readings 
are inaccurate.

1.  Verify correct ISE sensor type selection 

in the System Setup menu.

2.  Verify temperature readings are accurate.
3.  Recalibrate the pH or ORP sensor.
4.  Clean the pH or ORP sensor.
5.  Contact YSI Tech Support.

pH values display 
Over or Undr on 
Run screen. 

1.  Verify correct ISE sensor type selection 

in the System Setup menu. 

2.  Verify temperature readings are accurate.
3.  Recalibrate the pH sensor. 
4.  Clean the pH sensor and recalibrate.
5.  Contact YSI Tech Support.

ORP values display 
Over or Undr on 
Run screen. 

1.  Verify correct sensor type selection 

in the System Setup menu.

2.  Sample ORP value is outside the 

measurement range of -1500 to 1500 mV. 

3.  Verify temperature readings are accurate.
4.  Recalibrate the ORP sensor. 
5.  Clean the ORP sensor and recalibrate.
6.  Contact YSI Tech Support.

S P E C I F I C A T I O N S

These specifications represent typical performance and are subject to 
change without notice.  For the latest product specification information, 
please visit YSI’s website at ysi.com or contact YSI Tech Support. 

Parameter

Range

Resolution

Accuracy

Temperature

-5 to 55°C*

0.1°C

± 0.2°C

pH

0 to 14 
pH units

0.01 

Instrument with cable 
and sensor: +/- 0.2

ORP

-1500 to 
1500 mV

1 mV

Instrument with cable 
and sensor: +/-20 mV

A C C E S S O R I E S   /   P A R T   N U M B E R S

Part Number

Description

6050010

Pro10 Instrument

62610-1,-4, 
-10,-20, or-30

1, 4, 10, 20, 30-meter cable assembly*
(3.2, 13, 32.8, 65.6, 98.4-feet)

605101

pH Sensor

605323

Amplified pH Sensor

605102

ORP Sensor

603075

Carrying case, large, soft-sided

603162

Carrying case, small, soft-sided (holds instrument and up to 
a 4 meter cable/probe assembly)

603074

Carrying case, hard-sided

603069

Belt clip for clipping instrument onto belt

063517

Ultra clamp for clamping instrument to lab counter or other 
surface

063507

Tripod for instrument

603062

Cable management kit, included with all cables longer than 
1 meter

605978

Cable weight, 4.9 oz, stackable

603070

Shoulder strap

003821

pH 4 Buffer, box of 6 pints

003822

pH 7 Buffer, box of 6 pints

003823

pH 10 Buffer, box of 6 pints

603824

pH Buffer, assorted case, 2 pints each of buffer 4, 7 and 10

*All cables include a temperature sensor. The pH and ORP sensors are 
sold separately.  

Содержание Pro10

Страница 1: ...USER MANUAL English Pro10 User Manual DOCUMENT 605192 ...

Страница 2: ...ttery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Run Screen 5 Backlight 6 Powering Off 6 Navigation 6 First Power On 7 System Setup Menu 7 Audio 8 Contrast 8 ISE Sensor Type 8 ISE Units 9 ISE Auto Stable 9 pH Buffer Set 10 Language 10 Temperature Units 10 Auto Shutoff 10 Resetting the System Setup Menu to Factory Default 11 Calibration 11 Temperature 11 pH Calibration 12 ORP Calibr...

Страница 3: ...ct YSI Customer Service in Yellow Springs Ohio at 1 937 767 7241 800 897 4151 or visit YSI com for a Product Return Form Send the product and proof of purchase transportation prepaid to the Authorized Service Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the original warranty ...

Страница 4: ...r missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased Battery Installation The instrument requires 2 alkaline C cell batteries Under normal conditions the average battery life is 425 hours at room temperature without using the back light A battery symbol will blink in the lower left corner of the display to in...

Страница 5: ...h box options along the bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Press and hold for 3 seconds to turn instrument off 4 Menu Press to enter the System Setup menu from the Run screen 5 Enter Press to confirm entries and selections 6 Down Arrow Use ...

Страница 6: ...5 Push the sensor into the connector on the cable until it is firmly seated with only 1 o ring visible Failure to properly seat the sensor may result in damage 6 Twist the sensor clockwise to engage the threads and finger tighten Do NOT use a tool This connection is water tight The sensor is shipped with the tip in a storage bottle To remove twist the bottle off the lid and remove the bottle from ...

Страница 7: ...u the up and down arrow keys will move the highlighted bar up and down the system setup options See the System Setup menu section of this manual for more information about these options First Power On The instrument will step through an initial configuration when powered on for the first time to select the language Use the up or down arrow keys to highlight the appropriate language then press ente...

Страница 8: ...to open a submenu that allows you to select the ISE units to be displayed on the Run screen Highlight a unit and press enter to enable or disable it An enabled ISE unit will have an X in the box next to it Highlight the ESC Exit box along the bottom of the display and press enter to save any changes and to close the ISE Units submenu When pH is enabled in the ISE Sensor Type menu there are two sel...

Страница 9: ...s and to close the Temperature Units submenu Auto Shutoff Auto Shutoff allows you to set the instrument to turn off automatically after a period of time since the last button press In the setup menu use the up or down arrow keys to highlight Auto Shutoff then press enter to open the submenu Press enter while the minute field is highlighted to make it adjustable Next use the up or down arrow keys t...

Страница 10: ...g to stabilize Note the pH mV reading pH mVs in buffer 4 should be 159 to 180 mV from the previous buffer 7 pH mV value pH mVs in buffer 10 should be 159 to 180 mV from the previous buffer 7 pH mV value 10 Press enter to complete the calibration or press Cal to cancel 11 Calibration Successful will display for a few seconds to indicate a successful calibration and then the instrument will return t...

Страница 11: ...s Before taking measurements be sure the instrument has been calibrated to ensure the most accurate readings Install the sensor guard to protect the sensor Place the probe in the sample to be measured and give the probe a quick shake to release any air bubbles Be sure the sensors are completely submerged in the sample pH ORP pH and ORP readings are typically quick and accurate However it may take ...

Страница 12: ...tial For example if there are 15 records and number 12 is erased then 13 becomes 12 14 becomes 13 and 15 becomes 14 The instrument will return to Data mode after erasing one data set Select Erase All Data Sets and press enter to clear the Pro10 memory and return to Data mode Exiting Data Mode While in Data mode press enter to access the bottom functions Next highlight the ESC Exit box and press en...

Страница 13: ...ring to diminish potentially causing leaks If excess grease is present remove it using a lens cloth or lint free cloth General Maintenance Sensor Port It is important that the entire sensor connector end be dry when installing removing or replacing a sensor This will prevent water from entering the port Once the sensor is removed examine the i i connector inside the port If any moisture is present...

Страница 14: ...he sensor with clean water not DI water and then soak for at least 1 hour in clean water with occasional stirring to remove residual bleach from the junction If possible soak the sensor for a period of time longer than 1 hour in order to be certain all traces of chlorine bleach are removed Then rerinse the sensor with clean water and retest i Dry the port and sensor connector with compressed air a...

Страница 15: ...f 1500 to 1500 mV 3 Verify temperature readings are accurate 4 Recalibrate the ORP sensor 5 Clean the ORP sensor and recalibrate 6 Contact YSI Tech Support S pecifications These specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Support Parameter Range Resol...

Страница 16: ...ommitted to reducing the environmental footprint in the course of doing business Even though materials reduction is the ultimate goal we know there must be a concerted effort to responsibly deal with materials after they ve served a long productive life cycle YSI s recycling program ensures old equipment is processed in an environmentally friendly way reducing the amount of materials going to land...

Страница 17: ...authorized service centers throughout the United States and Internationally For the nearest service center information please visit ysi com and click Support or contact YSI Technical Support directly at 800 897 4151 1 937 767 7241 When returning a product for service include the Product Return form with cleaning certification The form must be completely filled out for a YSI Service Center to accep...

Страница 18: ...Item 605192 Rev A July 2014 2014 YSI Incorporated ...

Отзывы: