contamination (i.e. debris, grit, etc.) and cleaned with water and mild 
detergent if necessary.  

The same inspection should be made of the o-rings associated with 
the pH or ORP sensor connector when replacing a sensor. The o-rings 
should be free of dirt or debris before installing the sensor onto the 
cable. If no dirt or damage to the o-rings is evident, then they should 
be lightly greased with the o-ring grease provided without removing 
them from their groove. However, if there is any indication of damage, 
the sensor o-ring should be replaced with an identical o-ring. 

To remove the sensor o-rings:
Use a small, flat-bladed screwdriver or similar blunt-tipped tool to 
remove the o-ring from its groove near the sensor connector. Check 
the o-ring and the groove for any excess grease or contamination. If 
contamination is evident, clean the o-ring and nearby plastic parts with 
lens cleaning tissue or equivalent lint-free cloth. Alcohol can be used 
to clean the plastic parts, but use only water and mild detergent on the 
o-ring itself. Also, inspect the o-rings for nicks and imperfections. 

Using alcohol on o-rings may cause a loss of elasticity and may 
promote cracking. Do not use a sharp object to remove the 
o-rings.  Damage to the o-ring or the groove may result.

To re-install the sensor o-rings:
Place a small amount of o-ring grease between your thumb and index 
finger.  (More grease is not better!) Draw the o-ring through the grease 
while pressing the fingers together to place a very light covering of 
grease on the o-ring. Place the o-ring into its groove ensuring it does 
not twist or roll. Use the previously grease-coated finger to once again 
lightly go over the surface of the o-ring.

Do not over-grease the o-rings. Excess grease may collect 
particles that can compromise the seal. Excess grease can also 
cause the waterproofing capabilities of the o-ring to diminish, 
potentially causing leaks. If excess grease is present, remove it 
using a lens cloth or lint-free cloth.

GENERAL MAINTENANCE - SENSOR PORT

It is important that the entire sensor connector end be dry when 
installing, removing or replacing a sensor. This will prevent water 
from entering the port. Once the sensor is removed, examine the 

i

i

connector inside the port. If any moisture is present, use compressed 
air to completely dry the connector or let it air dry. If the connector 
is corroded, contact YSI Technical Support or the YSI authorized 
distributor where you purchased the instrument.  

Remove sensors upside down (facing the ground) to help 
prevent water from entering the port upon removal.

sensor MaIntenance

Typical working life for pH and ORP sensors is approximately 
12-24 months depending on usage, storage and maintenance. 
Proper storage and maintenance generally extends the sensor’s 
working life. 

SENSOR MAINTENANCE - TEMPERATURE

You must keep the temperature sensor free of build up. No additional 
maintenance is required. A toothbrush can be used to scrub the 
temperature sensor if needed.
 

SENSOR MAINTENANCE - pH AND ORP

caUtIon

: When using a cotton swab, be careful NOT to wedge 

the swab between the guard and the glass sensor. If necessary, 
remove cotton from the swab tip, so the cotton can reach all 
parts of the sensor tip without stress.  You can also use a pipe 
cleaner for this operation if more convenient. 

Cleaning is required whenever deposits or contaminants appear on 
the glass or platinum sensor surfaces or when the sensor’s response 
slows.  The cleaning can be chemical and/or mechanical.

Removing the sensor from the cable may make cleaning easier.   Initially, 
use clean water and a soft clean cloth, lens cleaning tissue, or cotton 
swab to remove all foreign material from the glass bulb or platinum 
button. Then use a moistened cotton swab to carefully remove any 
material that may be blocking the reference electrode junction of the 
sensor.

If good pH or ORP response is not restored, perform the following 
additional procedure:

i

i

i

Содержание Pro10

Страница 1: ...USER MANUAL English Pro10 User Manual DOCUMENT 605192 ...

Страница 2: ...ttery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Run Screen 5 Backlight 6 Powering Off 6 Navigation 6 First Power On 7 System Setup Menu 7 Audio 8 Contrast 8 ISE Sensor Type 8 ISE Units 9 ISE Auto Stable 9 pH Buffer Set 10 Language 10 Temperature Units 10 Auto Shutoff 10 Resetting the System Setup Menu to Factory Default 11 Calibration 11 Temperature 11 pH Calibration 12 ORP Calibr...

Страница 3: ...ct YSI Customer Service in Yellow Springs Ohio at 1 937 767 7241 800 897 4151 or visit YSI com for a Product Return Form Send the product and proof of purchase transportation prepaid to the Authorized Service Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the original warranty ...

Страница 4: ...r missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased Battery Installation The instrument requires 2 alkaline C cell batteries Under normal conditions the average battery life is 425 hours at room temperature without using the back light A battery symbol will blink in the lower left corner of the display to in...

Страница 5: ...h box options along the bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Press and hold for 3 seconds to turn instrument off 4 Menu Press to enter the System Setup menu from the Run screen 5 Enter Press to confirm entries and selections 6 Down Arrow Use ...

Страница 6: ...5 Push the sensor into the connector on the cable until it is firmly seated with only 1 o ring visible Failure to properly seat the sensor may result in damage 6 Twist the sensor clockwise to engage the threads and finger tighten Do NOT use a tool This connection is water tight The sensor is shipped with the tip in a storage bottle To remove twist the bottle off the lid and remove the bottle from ...

Страница 7: ...u the up and down arrow keys will move the highlighted bar up and down the system setup options See the System Setup menu section of this manual for more information about these options First Power On The instrument will step through an initial configuration when powered on for the first time to select the language Use the up or down arrow keys to highlight the appropriate language then press ente...

Страница 8: ...to open a submenu that allows you to select the ISE units to be displayed on the Run screen Highlight a unit and press enter to enable or disable it An enabled ISE unit will have an X in the box next to it Highlight the ESC Exit box along the bottom of the display and press enter to save any changes and to close the ISE Units submenu When pH is enabled in the ISE Sensor Type menu there are two sel...

Страница 9: ...s and to close the Temperature Units submenu Auto Shutoff Auto Shutoff allows you to set the instrument to turn off automatically after a period of time since the last button press In the setup menu use the up or down arrow keys to highlight Auto Shutoff then press enter to open the submenu Press enter while the minute field is highlighted to make it adjustable Next use the up or down arrow keys t...

Страница 10: ...g to stabilize Note the pH mV reading pH mVs in buffer 4 should be 159 to 180 mV from the previous buffer 7 pH mV value pH mVs in buffer 10 should be 159 to 180 mV from the previous buffer 7 pH mV value 10 Press enter to complete the calibration or press Cal to cancel 11 Calibration Successful will display for a few seconds to indicate a successful calibration and then the instrument will return t...

Страница 11: ...s Before taking measurements be sure the instrument has been calibrated to ensure the most accurate readings Install the sensor guard to protect the sensor Place the probe in the sample to be measured and give the probe a quick shake to release any air bubbles Be sure the sensors are completely submerged in the sample pH ORP pH and ORP readings are typically quick and accurate However it may take ...

Страница 12: ...tial For example if there are 15 records and number 12 is erased then 13 becomes 12 14 becomes 13 and 15 becomes 14 The instrument will return to Data mode after erasing one data set Select Erase All Data Sets and press enter to clear the Pro10 memory and return to Data mode Exiting Data Mode While in Data mode press enter to access the bottom functions Next highlight the ESC Exit box and press en...

Страница 13: ...ring to diminish potentially causing leaks If excess grease is present remove it using a lens cloth or lint free cloth General Maintenance Sensor Port It is important that the entire sensor connector end be dry when installing removing or replacing a sensor This will prevent water from entering the port Once the sensor is removed examine the i i connector inside the port If any moisture is present...

Страница 14: ...he sensor with clean water not DI water and then soak for at least 1 hour in clean water with occasional stirring to remove residual bleach from the junction If possible soak the sensor for a period of time longer than 1 hour in order to be certain all traces of chlorine bleach are removed Then rerinse the sensor with clean water and retest i Dry the port and sensor connector with compressed air a...

Страница 15: ...f 1500 to 1500 mV 3 Verify temperature readings are accurate 4 Recalibrate the ORP sensor 5 Clean the ORP sensor and recalibrate 6 Contact YSI Tech Support S pecifications These specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Support Parameter Range Resol...

Страница 16: ...ommitted to reducing the environmental footprint in the course of doing business Even though materials reduction is the ultimate goal we know there must be a concerted effort to responsibly deal with materials after they ve served a long productive life cycle YSI s recycling program ensures old equipment is processed in an environmentally friendly way reducing the amount of materials going to land...

Страница 17: ...authorized service centers throughout the United States and Internationally For the nearest service center information please visit ysi com and click Support or contact YSI Technical Support directly at 800 897 4151 1 937 767 7241 When returning a product for service include the Product Return form with cleaning certification The form must be completely filled out for a YSI Service Center to accep...

Страница 18: ...Item 605192 Rev A July 2014 2014 YSI Incorporated ...

Отзывы: