ADA-WA 1; ADA-WA 4
Installation recommendations
7
ba76071e01
02/2012
The marking on the sensor shaft points towards the
pipeline. On page 9 you will find a description of how to
correctly align the sensor with the aid of a marking aid on
the sensor plug head connector.
90 ° pipe installation
Note
The 90 ° pipe installation is only suitable for large pipe
diameters and high turbidity/total suspended solids values
respectively. For other cases, the 45 °installation should be
preferred.
For installation, select a location with a pipe diameter as
big as possible.
Rotate the sensor so that the marking on the shaft (arrow
symbol) points in the direction of the pipe axis. On page 9
you will find a description of how to correctly align the
sensor with the aid of a marking aid on the sensor plug
head connector.
ESS/WA...
WA 700/...
Marking aid in
this direction
Large ballcock
turned
ADA-WA1