Installation recommendations
ADA-WA 1;
AD
A-W
A
4
4
ba7
6071e01
02/
2012
Overview of the recommended installation locations
Note
Always follow the I
NSTALLATION
chapter of the respective sensor operating manual.
Sensor type
Installation
location
(see
diagram on
page 3)
Advantages (+) /
Disadvantages (-)
Conclusion / Comments
TriOxmatic 700 IQ,
TriOxmatic 701 IQ,
TriOxmatic 702 IQ
FDO 700 IQ
B
+ Good flow
+ Low susceptibility to contamina-
tion
+ Low risk of membrane damage
TetraCon 700 IQ
B
+ Good flow
+ Low susceptibility to contamina-
tion
Check and adjust cell constants
after installation (see sensor op-
erating manual)
SensoLyt 700 IQ
C
+ Good flow
+ Low susceptibility to contamina-
tion
- Danger of small stones damaging
the electrode
The enclosed fixing ring instead
of the protective hood improves
the incident flow.
Attention: It is no protection
against small stones!
The best incident flow is
achieved when the diaphragm
points against the flow direction.
TFK 700 IQ
C
+ Good flow
+ Low susceptibility to contamina-
tion
The enclosed fixing ring instead
of the protective hood improves
the incident flow.