Installation
ADA-WA 1; ADA-WA 4
10
ba76071e01
02/2012
3
Insert the sensor with the
retaining ring in the
receiving tube (4) up to
the stop.
4
Screw the sensor
adapter (5) on the unit.
The sensor is ready to be
installed in the
retractable armature.
5
Install the sensor in the
retractable armature (see
operating manual of the
retractable armature).
6
Connect the SACIQ
sensor connection cable
to the plug head
connector of the sensor.
7
If necessary, rotate the
sensor by means of the
marking aid until it is in
the correct position.
* The receiving tube for installation with the ADA-WA 1 is
included in the scope of delivery of the retractable
armature. For installation with the ADA-WA 4, use the
special (longer) receiving tube provided with the adapter in
order to ensure the correct length compensation.
ADA-W
A1
ADA-W
A1
4 *
5