Communications
Section 7
YSI Incorporated
Environmental Monitoring Systems Operations Manual
7-
4
Manufacturer: YSIIWQSG
Model:
EMS600_
Version:
1.07
Master: 1M!
Sonde: 10617[CR][LF]
The bus master sent a measurement command. The Sonde will take a maximum of 61 seconds to finish the
measurement. Upon completion, it will have 7 sensor data available.
Sonde: 1[CR][LF]
Master: 1D0!
Sonde: 1+17.5+12.05+98.7+8.25+6.45[CR][LF]
Master: 1D1!
Sonde: 1-325+10[CR][LF]
After finishing the measurement, the Sonde sent a service request to indicate completion. The bus master
then sent the "D0" command to retrieve the data. There were 5 data returned. Since 7 readings should be
available, the master continued with "D1" command and received the remaining data. The responses from
"D0" and "D1" commands are:
Temperature:
17.5
Specific conductance:
12.05
DO %:
98.7
DO mg/L:
8.25
pH (ISE1):
6.45
ORP (ISE2):
-325
Depth:
10
Содержание 6 series
Страница 195: ...650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3 65 ...
Страница 196: ...650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3 66 ...
Страница 297: ...Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9 5 ...
Страница 305: ...Required Notice Appendix B YSI Incorporated Environmental Monitoring Systems Operations Manual B 2 ...
Страница 319: ...Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E 6 ...
Страница 327: ...EMC Performance Appendix H YSI Incorporated Environmental Monitoring Systems Operations Manual H 2 ...
Страница 337: ...Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I 10 ...
Страница 341: ...Percent Air Saturation Appendix J YSI Incorporated Environmental Monitoring Systems Operations Manual J 4 ...
Страница 345: ...PAR Sensor Appendix K YSI Incorporated Environmental Monitoring Systems Operations Manual K 4 ...
Страница 359: ...ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M 10 ...
Страница 363: ...NMEA Applications Appendix N YSI Incorporated Environmental Monitoring Systems Operations Manual N 4 ...