Sondes
Section 2
YSI Incorporated
Environmental Monitoring Systems Operations Manual
2-8
your fingers. Once the O-ring is in position, squeeze it every 90 degrees to equalize the tension. See
Figure 11 above. DO
NOT
USE GREASE OR LUBICANT OF ANY KIND ON THE O-RING.
Using a hobby knife or a scalpel, trim the excess Teflon from the membrane, making your cut about 1/8
inch below the O-ring as shown in Figure 12 below. A razor blade can be used for the cut if no knife or
scalpel is available.
Figure 12
Figure 13
If the installation has been done properly, the finished product should have no bubbles, wrinkles, or tears as
shown in Figure 13 above.
NOTE:
Observe the following cautions to assure that your membrane installation is proper:
Secure the sonde tightly so that it will not move during membrane installation.
Wash hands before installation and do not allow finger oils or O-ring lubricant to touch the probe face
or the membrane.
Use caution when replacing the probe guard that you do not touch the membrane. If you suspect that
the membrane has been damaged, replace it immediately.
2.3.2 STEP 2 - INSTALLING THE PROBES
Remove the calibration cup from your sonde by hand as shown in Figure 14, to expose the bulkhead.
Figure 14
TRANSPORT CUP
BULKHEAD WITH
PROBE PORT PLUGS
Содержание 6 series
Страница 195: ...650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3 65 ...
Страница 196: ...650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3 66 ...
Страница 297: ...Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9 5 ...
Страница 305: ...Required Notice Appendix B YSI Incorporated Environmental Monitoring Systems Operations Manual B 2 ...
Страница 319: ...Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E 6 ...
Страница 327: ...EMC Performance Appendix H YSI Incorporated Environmental Monitoring Systems Operations Manual H 2 ...
Страница 337: ...Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I 10 ...
Страница 341: ...Percent Air Saturation Appendix J YSI Incorporated Environmental Monitoring Systems Operations Manual J 4 ...
Страница 345: ...PAR Sensor Appendix K YSI Incorporated Environmental Monitoring Systems Operations Manual K 4 ...
Страница 359: ...ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M 10 ...
Страница 363: ...NMEA Applications Appendix N YSI Incorporated Environmental Monitoring Systems Operations Manual N 4 ...