Troubleshooting
Section 6
YSI Incorporated
Environmental Monitoring Systems Operations Manual
6-
4
Water in probe connector
Dry connector; reinstall probe
2.3.2
Probe has been damaged
Replace probe
9
Calibration solutions out of spec or
contaminated with other solution
Use new calibration solutions
C
ORP fails Zobell check
Take into account temperature
dependence of Zobell solution
readings
5.4
Internal failure
Return sonde for service
9
Depth unstable or
inaccurate
Depth sensor has not been zeroed
Follow depth zero procedure
2.6.1
Depth sensor access hole is
obstructed
Follow depth cleaning procedure
2.10.2
Depth sensor has been damaged
Return sonde for service
9
Internal failure
Return sonde for service
9
Conductivity unstable
or inaccurate. Error
messages appear during
calibration.
Conductivity improperly
calibrated.
Follow cal procedure
2.6.1
Conductivity probe requires
cleaning
Follow cleaning procedure
2.10.2
Conductivity probe damaged
Replace probe
9
Calibration solution out of spec or
contaminated
Use new calibration solution
C
Internal failure
Return sonde for service
8
Calibration solution or sample
does not cover entire sensor.
Immerse sensor fully.
2.6.1
Installed probe has no
reading
The sensor has been disabled
Enable sensor
2.9.7
Water in probe connector
Dry connector; reinstall probe
2.3.2
Probe has been damaged
Replace the 6560 probe
9
Report output improperly set up
Set up report output
2.9.6
Internal failure
Return sonde for service.
9
Temperature, unstable
or inaccurate
Water in connector
Dry connector; reinstall probe
2.3.2
Probe has been damaged
Replace the 6560 probe
9
Optical probe (turbidity,
chlorophyll, rhodamine
WT, BGA-PC, or BGA-
PE) unstable or
inaccurate. Error
messages appear during
calibration
Probe requires cleaning.
Follow probe cleaning procedure
2.10.2
Probe requires calibration
Follow cal procedures
2.6.1
Probe has been damaged
Replace probe
9
Water in probe connector
Dry connector; reinstall probe
2.3.2
Calibration solutions out of spec
Use new calibration solutions
C
Wiper is not turning or is not
synchronized.
Activate wiper. Assure rotation.
Make sure setscrew is tight.
2.10.2
Wiper is fouled or damaged.
Clean or replace wiper.
2.10.2
Internal failure.
Return probe for service.
9
Содержание 6 series
Страница 195: ...650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3 65 ...
Страница 196: ...650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3 66 ...
Страница 297: ...Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9 5 ...
Страница 305: ...Required Notice Appendix B YSI Incorporated Environmental Monitoring Systems Operations Manual B 2 ...
Страница 319: ...Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E 6 ...
Страница 327: ...EMC Performance Appendix H YSI Incorporated Environmental Monitoring Systems Operations Manual H 2 ...
Страница 337: ...Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I 10 ...
Страница 341: ...Percent Air Saturation Appendix J YSI Incorporated Environmental Monitoring Systems Operations Manual J 4 ...
Страница 345: ...PAR Sensor Appendix K YSI Incorporated Environmental Monitoring Systems Operations Manual K 4 ...
Страница 359: ...ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M 10 ...
Страница 363: ...NMEA Applications Appendix N YSI Incorporated Environmental Monitoring Systems Operations Manual N 4 ...