SECTION 4 SETUP
From Main mode, press the [CALIBRATE] soft-key to enter the calibration menu. Notice that
the dark bar in the top right corner of the display shows the current mode.
98.0 %
8.15
mg/L
Calibrate
0.0
ppt
740
mmHg
24.38°C 11:33
AM
AUTO CAL
DO CAL
SETUP
DIAGNOSTICS
Next, press the [SETUP] soft-key to enter the Setup menu. The bottom row of the display will
show the following soft-key setup selections.
8.15
*
mg
/L
24.38°C 11:33
AM
DISPLAY
REPORT
STABLE
SYSTEM
These four selections are used to change the instrument setup.
4.1 DISPLAY
SETUP
Press the [DISPLAY] soft-key to show the following menu.
8.15
*
mg
/L
24.38°C 11:33
AM
SIZE
UNITS
CONTRAST
LAST DIG.
The [SIZE] soft-key cycles among the two possible display arrangements for Main mode. One
selection shows as many parameters as possible, the other shows Dissolved Oxygen in large type
with temperature and time in small type.
Setup
Display