SECTION 7 TROUBLESHOOTING CHART
Error Messages
The instrument performs a Power On Self Test each time it is turned on. The following error
codes are provided to facilitate troubleshooting.
Code/Problem Error
Possible
Cause
Correction
Blank Display
Dead / low batteries or
defective external power
supply
Replace batteries / test
power supply
5011 adapter improperly
connected
Unplug 5011 from power
supply connector
Internal power supply
failure
Verify circuit operation using
DVM. Replace PCB.
E1:
Check Probe
Temperature under
range
Connector improperly
installed
Check probe connection
Intermittent connection in
cable or plug
Repair or replace
Probe thermistor defective.
Should measure 2252
Ω
@
25
o
C
Replace probe sensor assy.
Temperature
over
range
Electrical leakage in
connector or cable
Repair or replace
Probe thermistor defective.
Should measure 2252
Ω
@
25
o
C
Repair probe sensor assy.
E2:
Check
Membrane
High Background
current (DO)
Insufficient warm-up time
See Calibration procedure
Incorrect probe zeroing
procedure
See Zero Calibration
Probe needs servicing
See probe maintenance
instructions
Probe malfunction
Repair or replace
E3:
Check
Membrane
DO Low current
Insufficient electrolyte or
fouled membrane
Install new membrane and
KCl
Contaminated
electrodes
See probe maintenance
instructions
Membrane too thick
Try another membrane
High resistance in probe
connection
Repair or replace