Code/Problem Error
Possible
Cause
Correction
E4:
Check
Membrane
DO High Current
Membrane too thin
Try another membrane
Electrodes
need
resurfacing
Repair or replace
Internal leakage in
probe or cable
connector
Repair or replace
Short circuit in probe or
cable
Repair or replace
E5:
Check
Membrane
DO unstable
Membrane damaged
Replace membrane
Probe needs servicing
See probe instructions
Connector
improperly
installed
Check probe connection
Intermittent
connection
in cable or plug
Repair or replace
Probe malfunction
Repair or replace
ERROR:
DO reading < min. DO The DO reading at the start of the OUR or SOUR
measurement is below the “minimum beginning
DO” entered in the OUR or SOUR setup.
See
ERROR:
Illegal value
The value entered is
outside the acceptable
range.
Enter the correct value
ERROR:
RTC defect control
register
Real time clock failure
Return for repair
ERROR:
Init. Logger failed
Logger initialization
failed
Return for repair