5-48
IM 760151-17E
SYNChroniz e
:
ALL
<Space>
U
<x>
I
<x>
NONE
ELEMent
<x>
<Space>
U
<x>
I
<x>
NONE
?
?
NULL
<Space>
OFF
ON
<NRf>
?
?
?
;
POVer
:INPut?
Function
Queries all settings related to the input element.
Syntax
:INPut?
Example
:INPUT? -> :INPUT:
WIRING P1W2,P1W2,P1W2;VOLTAGE:
RANGE:ELEMENT1 1.0000E+03;
ELEMENT2 1.0000E+03;
ELEMENT3 1.0000E+03;
ELEMENT4 1.0000E+03;:INPUT:CURRENT:
TERMINAL:ELEMENT1 DIRECT;
ELEMENT2 DIRECT;ELEMENT3 DIRECT;
ELEMENT4 DIRECT;:INPUT:CURRENT:
RANGE:ELEMENT1 5.00E+00;
ELEMENT2 5.00E+00;
ELEMENT3 5.00E+00;
ELEMENT4 5.00E+00;:INPUT:CURRENT:
SRATIO:ELEMENT1 10.0000;
ELEMENT2 10.0000;ELEMENT3 10.0000;
ELEMENT4 10.0000;:INPUT:FILTER:
LINE:ELEMENT1 OFF;ELEMENT2 OFF;
ELEMENT3 OFF;ELEMENT4 OFF;
ELEMENT5 OFF;:INPUT:FILTER:ZCROSS:
ELEMENT1 OFF;ELEMENT2 OFF;
ELEMENT3 OFF;ELEMENT4 OFF;:INPUT:
SCALING:STATE:ELEMENT1 0;
ELEMENT2 0;ELEMENT3 0;ELEMENT4 0;
ELEMENT5 0;:INPUT:SCALING:PT:
ELEMENT1 1.0000;ELEMENT2 1.0000;
ELEMENT3 1.0000;ELEMENT4 1.0000;
ELEMENT5 1.0000;:INPUT:SCALING:CT:
ELEMENT1 1.0000;ELEMENT2 1.0000;
ELEMENT3 1.0000;ELEMENT4 1.0000;
ELEMENT5 1.0000;:INPUT:SCALING:
SFACTOR:ELEMENT1 1.0000;
ELEMENT2 1.0000;ELEMENT3 1.0000;
ELEMENT4 1.0000;:INPUT:SYNCHRONIZE:
ELEMENT1 I1;ELEMENT2 I2;
ELEMENT3 I3;ELEMENT4 I4:INPUT:
NULL 0
[:INPut]:CURRent?
Function
Queries all settings related to the current
measurement.
Syntax
[:INPut]:CURRent?
Example
:INPUT:CURRENT? -> :INPUT:CURRENT:
TERMINAL:ELEMENT1 DIRECT;
ELEMENT2 DIRECT;ELEMENT3 DIRECT;
ELEMENT4 DIRECT;:INPUT:CURRENT:
RANGE:ELEMENT1 5.00E+00;
ELEMENT2 5.00E+00;
ELEMENT3 5.00E+00;
ELEMENT4 5.00E+00;:INPUT:CURRENT:
SRATIO:ELEMENT1 10.0000;
ELEMENT2 10.0000;ELEMENT3 10.0000;
ELEMENT4 10.0000;
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
[:INPut]:CURRent:AUTO[:ALL]
Function
Collectively turns ON/OFF the current auto
range of all power measurement elements.
Syntax
[:INPut]:CURRent:AUTO
[:ALL] {<Boolean>}
Example
:INPUT:CURRENT:AUTO:ALL ON
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
5.10 INPut Group