5-67
IM 760151-17E
Commands
5
:NUMeric:IMPedance:ITEM<x>
Function
Sets the numeric data output items for power
measurement or queries the current setting.
Syntax
:NUMeric:IMPedance:ITEM<x>
{NONE|<Function>[,<Element>]}
:NUMeric:IMPedance:ITEM<x>?
<x> = 1 to 16 (item number)
NONE = No output item
<Function>
=
{BU|BI|BP|ZR|ZI|Z|PHI|U|
I|FREQ}
<Element>
= {
<NRf>
}(
<NRf>
= 1 to 5)
Example
:NUMERIC:IMPEDANCE:ITEM1 BU,4
:NUMERIC:IMPEDANCE:ITEM1? ->
:NUMERIC:IMPEDANCE:ITEM1 BU,4
Description • The details of the item specified by
<Function>
are indicated below.
BU = Voltage U (dc) of the battery power
measurement element (display: Battery
Voltage)
BI = Current I (dc) of the battery power
measurement element (display: Battery
Current)
BP = Power P (dc) of the battery power
measurement element (display: Battery
Power)
ZR = Real part of the circuit impedance
(display: Z’)
ZI = Imaginary part of the circuit impedance
(display: Z’’)
Z = Absolute value of the circuit impedance
(display:
|
Z
|
)
PHI = Phase difference (display:
Φ
)
U = Voltage of the impedance measurement
element (display: U)
I = Current of the impedance measurement
element (display: I)
FREQ = Frequency (display: Freq)
• If
<Function>
is set to FREQ,
<Element>
can be omitted.
:NUMeric:IMPedance:NUMber
Function
Sets the number of items of the numeric data
that is output by
“
:NUMeric:IMPedance:VALue?
” or queries
the current setting.
Syntax
:NUMeric:IMPedance:NUMber {<NRf>}
:NUMeric:IMPedance:NUMber?
<NRf>
= 1 to 16
Example
:NUMERIC:IMPEDANCE:NUMBER 6
:NUMERIC:IMPEDANCE:NUMBER? ->
:NUMERIC:IMPEDANCE:NUMBER 6
Description • “
:NUMeric:IMPedance:
VALue?
” outputs the numeric data from 1 to the
specified value in order according to this setting.
• By default, the number of items of numeric
data is set to “6.”
:NUMeric:IMPedance:PRESet
Function
Presets the output item pattern of numeric data
for impedance measurement.
Syntax
:NUMeric:IMPedance:PRESet {<NRf>}
<NRf>
= 1 to 4 (preset pattern number)
Example
:NUMERIC:IMPEDANCE:PRESET 1
Description • For details on the output items that are preset,
see “(2) Preset Pattern of Output Items of
Impedance Measurement Numeric Data.”
• By default, output items of “Pattern 2” is
selected.
:NUMeric:IMPedance:VALue?
Function
Queries the numeric data for impedance
measurement.
Syntax
:NUMeric:IMPedance:VALue?
Example
• Example in which “
:NUMeric:FORMat
” is set
to “
ASCii
”
:NUMERIC:IMPEDANCE:VALUE? ->
104.75E+00,105.02E+00,-
0.38E+00,..
(omitted)
..,49.868E+00
• Example in which “
:NUMeric:FORMat
” is set
to “
FLOat
”
:NUMERIC:IMPEDANCE:VALUE? ->
#4
(Number of bytes, 4 digits)(Series of data
bytes)
Description • Outputs the numeric data of items numbers in
order from 1 to
:NUMeric:IMPedance:NUMber
.
• For an array-type function
(
{ZR|ZI|Z|PHI|U|I|FREQ}
), numeric data
of the number of arrays are output in order
from 1 to
:NUMeric:IMPedance:ARRay
for
a single item.
• For the format of the individual numeric data
that is output, see “Numeric Data Format” at
the end of this group (section).
:NUMeric:NORMal?
Function
Queries all settings related to the numeric data
output for power measurement.
Syntax
:NUMeric:NORMal?
Example
:NUMERIC:NORMAL? -> :NUMERIC:
NORMAL:NUMBER 15;ITEM1 URMS,1;
ITEM2 UMN,1;ITEM3 UDC,1;
ITEM4 UAC,1;ITEM5 IRMS,1;
ITEM6 IMN,1;ITEM7 IDC,1;
ITEM8 IAC,1;ITEM9 P,1;ITEM10 S,1;
ITEM11 Q,1;ITEM12 LAMBDA,1;
ITEM13 PHI,1;ITEM14 FU,1;
ITEM15 FI,1
Description For the values of
“
:NUMeric[:NORMal]:ITEM<x>
,” the
numeric data output items for the amount
specified by “
:NUMeric[:NORMal]:
NUMber
” are output.
5.13 NUMeric Group