IM 11M10A01-03E
8-1
8. Troubleshooting
8.
Troubleshooting
The table below summarizes troubles that may develop in the OX102, problem source
and corrective action.
T801E.EPS
Abnormal symptom
Problem source
Corrective action
No indications
1. No power applied
1. Apply power
2. Fuse has blown
2. Replace the fuse
FAIL lamp lit
1. Sensor connector failure
1. Connect the sensor connector securely
2. Sensor heater is disconnected
2. Replace the sensor
Low value displayed
1. Operating pressure is out of specified range
1. Use within the specified pressure range
High value displayed
2. Improper calibration
2. Recalibration
3. Sensor failure
3. Replace the sensor
4. Expired activated carbon has been used
4. Replace the activated carbon filter
Slow response
1. Ambient pressure around sensor is negative
1. Use within the specified pressure range
2. Sensor has been clogged
2. Clean the sensor head
Large error in measurement
1. Operating temperature is out of specified range
1. Use within the specified temperature range
2. Operating pressure is out of specified range
2. Use within the specified pressure range
3. Sample gas contains oxidation-reduction
3. Specified accuracy cannot be expected in
gas or corrosive gas
measurement
4. Sensor head touches wall
4. Reinstall the sensor following procedure
as in Section 3.1
5. Expired activated carbon has been used
5. Replace the activated carbon filter
Occasional, large fluctuations
1. Sensor failure
1. Replace the sensor
in readings
2. Operating pressure is out of specified range
2. Use within the specified pressure range
Calibration fails
1. Improper calibration gas concentration
1. Use proper calibration gas
2. Sensor failure
2. Replace the sensor
3. Operating pressure is out of specified range
3. Recalibrate within the specified pressure
range
4. Useful life of activated charcoal filter is over 4. Replace activated charcoal filter
No alarm contact signal is
1. Alarm contact signal is not set
1. Set the alarm contact signal properly
activated1.
2. Alarm contact relay failure
2. Contact Yokogawa