IM 11M10A01-03E
6-24
ENT
ppm+CAL
ENT
ppm + CAL
ERROR
ENT
10000
F6319E.EPS
YES
NO
[10000 ppm O
2
Calibration]
Indication
Stop supplying calibration gas No.5
and supply No.6
(O
2
concentration allowance: 8500-9500 ppm)
Measurement mode
Entering calibration
value
Calibration value
Measured
value
Calculating
Changing
value
Calibration
completed
Note: To return to measurement mode, press [ppm + CAL]
%
s
ppm
d
m
.
%
s
ppm
d
%
s
ppm
d
%
s
ppm
d
(2) Moving zero point
This is to translate the output function: to change value b of the characteristic function, y
= ax + b. The range for moving the zero point is from -999 to +999 ppm.
F6320E.EPS
Operational Flow Chart
Measured
value
Increment/decrement
Calculating/
Configuring
Changing
value
Note: To return to measurement mode, press [ppm + CAL]
Indication
m
.
%
s
ppm
d
%
s
ppm
s
Figure 6.13 Flow Chart of Moving Zero Point