IM 11M10A01-03E
6-13
6. Operation
(5) Configuring high/low alarm contact output settings
This mode is to configure the oxygen concentration high/low alarm contact output
settingsalarm. Configure settings following the procedure shown in the flow chart
below. The high alarm setting should be equal or greater than the low alarm setting.
ENT
ppm+CAL
ENT
m
.
ppm+CAL
ALM
ALM
%
ppm
ALM
Yes
No
(
)
H.ALM
H.ALM
F6310E.EPS
%
d
ppm
s
Indication
Operational Flow Chart
Measured
Value
%
s
ppm
s
%
d
ppm
s
*
High alarm
To low alarm setting (following page)
(*) If the entered high alarm setting is less than the low alarm setting, the display will show
for about 3 seconds and the entered setting will flash to prompt reentry. An appropriate
value should be reentered.
High Alarm Setting
High Alarm
Low Alarm
>
=
Changing
value
Value
accepted
E- - -
Figure 6.8 Flow Chart of High/Low Alarm Setting