IM 05C01E22-10E
1-2
1.2
Notes on Setting Parameters
This section describes the setting parameters for using the communication functions and their setting
ranges.
NOTE
The details of UT150L communication functions need to be the same as those of the communication
functions of the host devices to be connected. Check the communication parameters of the host device
first, then set up those of the UT150L.
Table 1-1 Parameters to be Set for Communication Functions
Parameter Name
Stop bit
Address
Parity
PC link communication
Protocol selection
MODBUS communication
Baud rate
Note 1: When “3: ASCII mode” is selected for MODBUS communication in protocol selection, the data length is fixed to “7.”
When
“4: RTU mode” is selected, it is fixed to “8.”
Data length
Symbol
ADR
PSL
PRI
DLN
BPS
STP
1, 2
0: 2400, 1: 4800, 2: 9600
0: none, 1: even, 2: odd
Setting Range
1 to 99
7, 8 (Note 1)
0: without sum check
1: with sum check
3: ASCII mode
4: RTU mode
2: 9600
1
8
Default
1
0
1: EVN
●
Protocol-by-Protocol Default Parameter Settings
Note: Circled numbers denote fixed values.
PSL
BPS
PRI
STP
DLN
PC-link communication without sum check
0
9600
EVN
1
8
PC-link communication with sum check
1
9600
EVN
1
8
MODBUS communication (ASCII mode)
3
9600
EVN
1
7
MODBUS communication (RTU mode)
4
9600
EVN
1
8
Parameter
Communication Protocol
●
Protocol selection (PSL)
Set the same communication protocol as that of the host device to be connected. The UT150L has PC
link communication, and MODBUS communication functions.
Содержание UT150L
Страница 1: ...User s Manual Model UT150L Communication Functions IM 05C01E22 10E IM 05C01E22 10E 3rd Edition...
Страница 2: ......
Страница 8: ...IM 05C01E22 10E vi Appendix Table of ASCII Codes Alphanumeric Codes App 1 Revision Record i...
Страница 12: ...IM 05C01E22 10E 1 4...
Страница 14: ...IM 05C01E22 10E 2 2...
Страница 38: ...IM 05C01E22 10E 3 24...
Страница 48: ...IM 05C01E22 10E 4 10...
Страница 54: ...IM 05C01E22 10E 6 2...
Страница 56: ...IM 05C01E22 10E App 2...
Страница 58: ......
Страница 59: ......