< 2. Specifications >
2-2
IM 12A0V2-E
When option code “/RR” specified, two measurement range set in the FC400G can be switched
remotely.
Remote range switching contact input:
Applied voltage when contact input is opened: 5 V DC (Range 1)
Applied current when contact input is closed: 20 μA (Range 2)
Remote range switching contact output:
Contact type: dry contact output
Contact rating: 250 V AC, 1A, 125 VA resistance load)
220 V DC, 60 W (resistance load)
Contact status
of Power OFF
Contact status of Power ON
Open
Non-activated
Activated
Open
Closed
Open: Range 1, Closed: Range 2
-FC5, -PF7, -PF8 is equipped as standard with the remote range switching function for the
Non-reagent Type Free Available Chlorine Sensor Unit. Two measurement ranges can be
switched when -N2 on FLXA402T is specified. Three ranges can be switched when -N4 is
specified. For the specifications of the remote range switching contact input, remote range
switching contact output, refer to FC800D general specification GS 12F05B10-01EN.