< 2. Specifications >
2-1
IIM 12A0V2-E
2. SPECIFICATIONS
Measurement: Free available chlorine concentration of tap water, pH of tap water or industrial water.
Instruments integrated:
FC800D Non-reagent type Free Chlorine Sensor Unit /FLXA402T Liquid Analyzer for
Turbidity and Chlorine, or FC400G non-reagent type free available chlorine analyzer,
4-wire
pH
converter
system
(FLXA402)
Refer to respective General Specifications:
GS12F05B10-01EN:
FC800D/FLXA402T
GS
12F5A1-E:
FC400G
GS 12A01F01-01EN: FLXA402
GS
12B07B02-E:
PH8EFP
GS
12J05C02-00E:
PH8HF,
PUS400G
GS
19C01B05-01EN: PG400
Calibration method:
Chlorine analyzer: Manual/auto zero calibration
pH meter: Manual calibration
Cleaning system:
Chlorine analyzer: Not provided
pH analyzer: Ultrasonic cleaning
Sample (process) conditions:
Pressure:
100 to 750 kPa (Under 100 kPa, consult Yokogawa)
Flow rate:
Chlorine analyzer: 0.1 to 10 L/min
pH analyzer: 3 to 10 L/min
Temperature:
0 to 50 °C when using FC800D -F1 or -C1
0 to 40 °C when using FC800D -F2 or -F3
Note:
When the system is used for measurement in heavily contaminated water other than tap water and industrial
water, the filter of the pressure reducing valve may be clogged early. In that case take measures to avoid contam
-
ination, e.g., changing the sampling point.
Ambient conditions:
Temperature:
-5 to 50 °C
(Protection against freeze should be provided when necessary in winter)
Humidity:
5 to 95% RH, non-condensing (10 to 90 % for pH converter system)
Storage temperature: -30 to 70 °C
-25 to 70 °C when using PUS400G
Materials:
Piping:
Hard PVC, polypropylene resin
Stanchion: Carbon steel or stainless steel
Finish: Baked polyurethane resin coating
Color: Equivalent to Munsell 0.6GY3.1/2.0
Installation site: Indoors or in cabinet: not exposed to the weather. Minimal vibration and shock
allowed. Maintenance space required. No corrosive gases present. Moderate
humidity.
Piping connection:
Sample inlet:
VP16
Drain:
VP40
Cable inlet: Watertight plastic cable gland equivalent to JIS A15
Cable size: 9 to 12 mm OD
Power supply: 100 V AC ±10%, 50/60 Hz
110
V
AC
±10%,
50/60
Hz
200
V
AC
±10%,
50/60
Hz
220 V AC ±10%, 50/60 Hz (depending on system configuration)
Power consumption:
Approx.65 VA (with FC400G non-reagent type free available chlorine analyzer,
pH converter system, and Ultrasonic oscillator) (depending on system
configuration)
Apparent power of solenoid valve for auto zero calibration
When activated: 22/17 VA (50/60 Hz)
When non-activated: 35/27 VA (50/60 Hz)
Weight: Approx. 45 kg (depending on system configuration)