2-20
IM 701240-01E
B > Time Trigger, B < Time Trigger, and B TimeOut Trigger
A trigger is activated on the falling or rising edge of the pulse when the pulse width
(width over which condition B is met) exceeds (or drops below) a preset time. In the
case of a timeout trigger, a trigger is activated when a preset time elapses.
LogicA/LogicB
Disable: Do not use logic input
Enable: Use logic input
CH1 to CH16
L: Low
level
H: High
level
X: Don’t
care
• B > Time Trigger
B
Trigger
Time
B: Pulse width
• B < Time Trigger
B
Trigger
Time
• B TimeOut Trigger
B
Trigger
Time
2.3 Setting the Trigger