IM 77C01U01-01E

1st Edition : Dec. 26, 2000-00

 (Numeric up switch)

This switch is used for increment of setting
parameter and input/output adjustment data.

 (Numeric down switch)

This switch is used for decrement of setting
parameter and input/output adjustment data.

Note 1: Instantanous power value is displayed with symbol

in case only negative.

Note 2: When display of maximum value, ‘—’ is displayed at

the top with light on and off.

Note 3: When display of minimum value, ‘—’ is displayed at

the top with light on and off.

Note 4: Instantanous power factor is displayed only when

measuring option is designated.

Communication Output

RS-485 or LON communication outputs can optionally be

selected

<Communication data>
Following measuring value can be read out by converting input

into PT•CT primary side input.

• Integrated power
• Optional integrated power (present value)
• Optional integrated power (last value)
• Instantanous power
• Voltage r.m.s. value
• Current r.m.s. value
• Instantanous power factor or Current 2 r.m.s. value
• Voltage maximum value
• Valtage minimum value
• Current maximum value
• Current 2 maximum value
Also, start•stop of optional integrated power and reset of

maximum and minimum values of voltage r.m.s.
value and maximum value of current r.m.s. value
can be done through communication.

<RS-485 Communication>
Function: Various measuring values can be read out through

personal computer by command/ response mode.
Readout of measuring value would be made
individually or in block.

Also, control of optional integration and initialization of

maxium and mimimum values can be done
through persomal computer.

System Configuration:

Application Software

RS-232C Port

RS-232C/485 Conveter

RS-485

Personal 
Computer

Power
Monitor

Power
Monitor

Power
Monitor

Power
Monitor

Power
Monitor

Terminating

Note: RS-232C/485 Converter is recommended to use our

ML1 in AUTO mode.

Communication specs.: RS-485 Interface
Transmit distance: Maximum about 1.2km

(When use of 24AWG twist pair cable)

Optional integrated power:

䊐䊐䊐䊐䊐

[Wh] (w/o symbol, integer 5 digits)

Instantanous power:

±

䊐䊐䊐

.

[W] to 

±

䊐䊐䊐䊐

[MW] (w/symbol.

floating decimal point 4 digits, minimum
resolution:0.1W)

Voltage r.m.s. value:

䊐䊐䊐

.

[V] to 

䊐䊐䊐䊐

[kV] (w/o symbol, floating

decimal point 4 digits, minimum resolution:0.1V)

Current r.m.s. value:

.

䊐䊐䊐

[A] to 

.

䊐䊐䊐

[kA] (w/o symbol,

floating decimal point 4 digits, minimum
resolution:0.001A)

Voltage maximum value:

䊐䊐䊐

.

[V] to 

䊐䊐䊐䊐

[kV] (w/o symbol,

floating decimal point 4 digits, minimum
resolution:0.1V)

Voltage minimum value:

䊐䊐䊐䊐

[V] to —

䊐䊐䊐䊐

[kV] (w/o symbol,

floating decimal point 4 digits, minimum
resolution:0.1V)

Current maximum value:

.

䊐䊐䊐

[A] to 

.

䊐䊐䊐

[kA] (w/o symbol,

floating decimal point 4 digits, minimum
resolution:0.001A)

Instantanous power factor:

d

.

䊐䊐䊐

 to 1.000 to G

.

䊐䊐䊐

[COS

φ

]

(w/o symbol, fixed decimal point 4 digits,
minimum resolution:0.001COS

φ

, d:Lead, G:Lag)

kWh LED: Light on during display of integrated power

[kWh].

Wh LED:

Light on during display of optional integrated
power [Wh].

W LED:

Light on during display of instantanous power [w].

V LED:

Light on during display of voltage r.m.s. value [V].

A LED:

Light on during display of current r.m.s. value [A].

COS

φ

 LED: Light on during instantanous power factor [COS

φ

]

X10

3

 LED: Light on when displaying instantanous value is

kilo unit. Or lighton when integrated power [kWh]
is Mega unit.

X10

6

 LED: Light on when displaying instantanous value is

mega unit.

COMM LED: Green light on when RS-485 or LON communi-

cation. As for LON communication, red light on
and off when network parameter is under
construction, and red light on when communica-
tion trouble or service.

SET/ENT Switch: This switch changes-over display of

integrated power, optional integrated power,
instantanous power, voltage r.m.s. value, current
r.m.s. value and instantanous power factor. Also, it
selects parameter setting item and input/output
adjust item.

R, S, T (phase indicator)
Light on phase that the data is displayed in data display.
Current display: R, S, T
Voltage display: R-S, S-T, T-R

R, S, T (3-phase 4-wire system)

 (Range switch)

This switch is used for display line change-over,
and for movement column position of setting data
and decimal point position.

Содержание Powercert UZ005

Страница 1: ...User s Manual IM 77C01U01 01E Model UZ005 Power Monitor IM 77C01U01 01E 1st Edition ...

Страница 2: ...Blank Page ...

Страница 3: ...to connect the grounding 3 Turn off the product before installation and wiring 4 Use the power supply within specifications 5 Observe the following instructions for correct installation Recommended panel thickness is 1 to 10mm Install the product horizontally Install the product using attached fixture from the backside of the product Turn the screw only up to 180 after the screw touches the mounti...

Страница 4: ...t 1 2 110V can also be used at 120V input Note 3 Rated input 5 6 440V is the specs only for 1 phase 2 wire system and 3 phase 3 wire system Note 4 Rated input A to F is the specs for only 3 phase 4 wire system Rated input voltage indicates phase voltage A B 64V C D 127V E F 227V indicate 110V 3 220V 3 480V 3 respectively In case of 440V 3 input it can be used at E F 227V of rated input Note 5 Spec...

Страница 5: ...c 26 2000 00 External Dimensions and Panel Cutout Dimensions External Dimensions 96 90 8 90 8 96 11 100 30 Fixturs 91 8 112 Fixturs Panel thickness 1 to10 mm Terminal Cover Panel Cut Dimension Min 130 Min 117 92 0 8 0 92 0 8 0 Unit mm ...

Страница 6: ... to overtighten them Large bracket top mounting hardware Terminal board Small bracket bottom mounting hardware Panel Insert a screwdriver into the brackets to tighten the screws Direction to insert the product Insert the product into the opening at the front of the panel Terminal Arrangement 40 screw M3 5 1 phase 2 wire system Terminat No 1 2 51 52 3 4 53 54 5 6 55 56 9 59 10 60 31 32 33 34 35 36 ...

Страница 7: ...R 7 57 8 58 2 52 4 54 P1 P3 P2 9 10 1 51 3 53 5 55 6 56 PT CT CT CT K L K L K L 1S 1L 2S 2L 3S 3L P0 60 59 N 3 11 12 13 14 15 18 40 20 31 32 33 34 35 36 16 37 17 38 39 19 Power Side Load Side L N Power Supply Communication Short Terminating Voltage or Contact A B C Shild Receiver Device 2 1 T S R 3 phase 3 wire system 3 phase 4 wire system CAUTION The following caution for safety should be taken f...

Страница 8: ...ms RS 485 station number data transmission rate lower limit of input range upper limit of input range PT ratio CT ratio in tegrating low cut power integrating pulse unit fixed point part exponent integrating pulse ON pulse duration 3 Adjustment items power momentary value in put zero power momentary value input span voltage momentary value input zero volt age momentary input span current momen tar...

Страница 9: ...rated pulse unit characteristic setting 4 Integrated pulse ON width setting 4 RS 485 communication protocol setting 4 RS 485 station number setting 4 RS 485 data baud rate setting 4 LON communication node number setting 4 LON communication integrated power data transmission period setting minute 4 LON communication integrated power data transmission period setting second 4 LON communication power ...

Страница 10: ...nge numeral column at pressing key 5 Change numeral column at pressing and key 6 Fixing numeral value at pressing key 7 Light on and off of setting value 8 If the setting is complete correctly press key If required correcting press any of key key and key then repeat from step 1 7 Example of Communication Connection RS232 C Converter RS232 C Cable Short circuit terminating 33 34 35 33 34 35 36 37 A...

Страница 11: ...0000 MW and the value calculated by the formula below is within above input measuring range list Input range W Primary side input power W PT ratio x CT ratio Rated power factor LEAD 0 5 to 1 to LAG 0 5 Optional integrated control signal Input point 1 point Input signal contact or voltage Contact signal ON signal OFF signal below 200 over 100k Voltage signal 1V DC below 200 4 5 to 25V DC ON signal ...

Страница 12: ...9MW 10MW over Display decimal point 0 to 99999kWh 0 00 to 999 99MWh kWh 103 0 0 to 9999 9MWh kWh 103 0 to 99999kWh kWh 103 Inregrated power data is reset to zero when the data exceed maximunm display value Setting range Rated power 100W 200W 300W 400W 500W 600W 800W 1000W 1500W 2000W 3000W 4000W 5 556 10 6 to 1 000 10 1kWh pulse 6 667 10 6 to 1 000 10 1kWh pulse 1 000 10 5 to 1 000 10 1kWh pulse 1...

Страница 13: ...pair cable Optional integrated power 䊐䊐䊐䊐䊐 Wh w o symbol integer 5 digits Instantanous power 䊐䊐䊐 䊐 W to 䊐䊐䊐䊐 MW w symbol floating decimal point 4 digits minimum resolution 0 1W Voltage r m s value 䊐䊐䊐 䊐 V to 䊐䊐䊐䊐 kV w o symbol floating decimal point 4 digits minimum resolution 0 1V Current r m s value 䊐 䊐䊐䊐 A to 䊐 䊐䊐䊐 kA w o symbol floating decimal point 4 digits minimum resolution 0 001A Voltage ...

Страница 14: ...200 Data format Start bit 1 bit Data bit 8 bits Parity None Stop bit 1 bit Error detect SUM CHECK simply adding 2 bytes Xon Xoff Control None Terminating character designation Yes CR Station number setting Can be set through parameter setting screen Setting range 1 to 31 Communication error disposal If data received is other than command reading is ignored and no disposal be made noise or erroneou...

Отзывы: